Flow Cytometry Principles practice of Fluorescence Spectroscopy in
- Slides: 44
Flow Cytometry Principles & practice of “Fluorescence Spectroscopy in Biological Diagnosis & Research” Dr. Hekmatimoghaddam Assistant professor of pathology
Definitions n Flow Cytometry – Measuring properties of cells in flow n Flow Sorting – Sorting (separating) cells based on properties measured in flow – Also called Fluorescence-Activated Cell Sorting (FACS)
Basics of Flow Cytometry Fluidics • Cells in suspension • flow in single-file through • an illuminated volume where they Optics • scatter light and emit fluorescence • that is collected, filtered and Electronics • converted to digital values • that are stored on a computer
Fluidics n Need to have cells in suspension flow in single file through an illuminated volume n In most instruments, accomplished by injecting sample into a sheath fluid as it passes through a small (50 -300 µm) orifice
Flow Cell Injector Tip Fluorescence signals Focused laser beam Sheath fluid
Fluidics n When conditions are right, sample fluid flows in a central core that does not mix with the sheath fluid n This is termed Laminar flow
Fluidics n The introduction of a large volume into a small volume in such a way that it becomes “focused” along an axis is called Hydrodynamic Focusing
Fluidics - Differential Pressure System n Use air (or other gas) to pressurize sample and sheath containers n Use pressure regulators to control pressure on each container separately
Fluidics - Differential Pressure System n Sheath pressure will set the sheath volume flow rate (assuming sample flow is negligible) n Difference in pressure between sample and sheath will control sample volume flow rate n Control is not absolute - changes in friction cause changes in sample volume flow rate
Fluidics - Differential Pressure System C. Göttlinger, B. Mechtold, and A. Radbruch
Fluidics - Particle Orientation and Deformation “a: Native human erythrocytes near the margin of the core stream of a short tube (orifice). The cells are uniformly oriented and elongated by the hydrodynamic forces of the inlet flow. b: In the turbulent flow near the tube wall, the cells are deformed and disoriented in a very individual way. v>3 m/s. ” V. Kachel, et al. - MLM Chapt. 3
Fluidics - Flow Chambers Flow through cuvette (sense in quartz) H. B. Steen - MLM Chapt. 2
Flow Cell Injector Tip Fluorescence signals Focused laser beam Sheath fluid
Optics n Need to have a light source focused on the same point where cells have been focused (the illumination volume) n Two types of light sources – Lasers – Arc-lamps
Optics - Light Sources n Lasers – can provide a single wavelength of light (a laser line) or (more rarely) a mixture of wavelengths – can provide from milliwatts to watts of light – can be inexpensive, air-cooled units or expensive, water-cooled units – provide coherent light
Optics - Light Sources n Arc-lamps – provide mixture of wavelengths that must be filtered to select desired wavelengths – provide milliwatts of light – inexpensive, air-cooled units – provide incoherent light
Optics - Forward Scatter Channel n When a laser light source is used, the amount of light scattered in the forward direction (along the same axis that the laser light is traveling) is detected in the forward scatter channel n The intensity of forward scatter is proportional to the size, shape and optical homogeneity of cells (or other particles)
Forward Angle Light Scatter Laser FALS Sensor
Optics - Side Scatter Channel n When a laser light source is used, the amount of light scattered to the side (perpendicular to the axis that the laser light is traveling) is detected in the side or 90 o scatter channel n The intensity of side scatter is proportional to the internal structure and granularity of cells (or other particles)
90 Degree Light Scatter Laser FALS Sensor 90 LS Sensor
Optics - Light Scatter n Forward scatter tends to be more sensitive to surface properties of particles (e. g. , cell ruffling) than side scatter – can be used to distinguish live from dead cells n Side scatter tends to be more sensitive to inclusions within cells than forward scatter – can be used to distinguish granulated cells from non-granulated cells
Fluorescence Detectors Laser FALS Sensor Fluorescence detector (PMT 3, PMT 4 etc. )
Optics - Filter Properties n Long pass filters transmit wavelengths above a cut-on wavelength n Short pass filters transmit wavelengths below a cut-off wavelength n Band pass filters transmit wavelengths in a narrow range around a specified wavelength – Band width can be specified
Standard Long Pass Filters Light Source 520 nm Long Pass Filter Transmitted Light >520 nm Light Standard Short Pass Filters Light Source 575 nm Short Pass Filter Transmitted Light <575 nm Light
Standard Band Pass Filters 630 nm Band. Pass Filter White Light Source Transmitted Light 620 -640 nm Light
Optics - Filter Properties n When a filter is placed at a 45 o angle to a light source, light which would have been transmitted by that filter is still transmitted but light that would have been blocked is reflected (at a 90 o angle) n Used this way, a filter is called a dichroic filter or dichroic mirror
Dichroic Filter/Mirror Filter placed at 45 o Light Source Transmitted Light Reflected light
Optics - Filter Layout n To simultaneously measure more than one scatter or fluorescence from each cell, we typically use multiple channels (multiple detectors) n Design of multiple channel layout must consider – spectral properties of fluorochromes being used – proper order of filters and mirrors
Common Laser Lines 350 300 nm 457 488 514 400 nm 500 nm 610 632 600 nm 700 nm PE-TR Conj. Texas Red PI Ethidium PE FITC cis-Parinaric acid
Example Channel Layout for PMT Laser-based Flow Cytometry 4 Flow cell PMT Dichroic Filters 3 PMT 2 Bandpass Filters PMT 1 Laser
Optics - Detectors n Two common detector types – Photodiode § used for strong signals when saturation is a potential problem (e. g. , forward scatter detector) – Photomultiplier tube (PMT) § more sensitive than photodiode but can be destroyed by exposure to too much light
Summary of Part 1 Fluidics • Cells in suspension • flow in single-file through • an illuminated volume where they Optics • scatter light and emit fluorescence • that is collected, filtered and Electronics • converted to digital values • that are stored on a computer
Typical Research Cytometer (Coulter 753) (1980 s) $200 -300, 000 Detectors Lasers Fluidics Computers Laser Power Supply
Typical Clinical Cytometer Computer System Detector & Mechanical $90 -120, 000 Fluidics
Clinical Applications Of Flow Cytometric Analysis Flow. Cytometric(immunophenotypic) Classification Of Leukemias
Immunophenotyping CD 4 CD 2
Immunophenotyping . 1 1 10 Log FITC 1000
CD 4/CD 8 Quadstats 1 2 45% 2% 27% 3 . 1 1 26% 10 100 Log FITC Fluorescence (CD 8) 4 1000
The Cell Cycle G 2 S M G 1 G 0 Quiescent cells
Normal Cell Cycle G 2 M G 0 DNA Analysis G 1 s G 0 G 1 C o u n t s 0 200 400 G 2 M 600 4 N 2 N DNA content 800 1000
DNA Analysis DNA index 1. 21 Aneuploid peak 0 200 400 600 PI Fluorescence 800 1000
Reticulocyte Analysis RMI = 34 RMI = 0 R 4 R 3 R 2 R 1 . 1 1 10 100 log Thiazole Orange R 4 R 3 R 2 R 1 1000 . 1 1 10 1000 log Thiazole Orange
Light Scatter Gating Side Scatter Projection Neutrophils Scale 1000 200 100 50 40 Monocytes 30 20 15 Lymphocytes 8 0 200 400 600 800 90 Degree Scatter 1000
- Principles of fluorescence spectroscopy
- Basic principles flow cytometry
- Principle of fluorescence spectroscopy
- Fluorescence spectroscopy - ppt
- Atomic fluorescence spectroscopy principle
- Explain quantum yield
- Purdue university flow cytometry
- Resolution impact matrix
- Olisambu uche
- Contrad flow cytometry
- Flow cytometry simulator
- Apd vs pmt flow cytometry
- Flow cytometry cell counting
- Long pass filter flow cytometry
- Boolean gating flow cytometry
- Einstein flow cytometry
- Einstein flow cytometry
- Einstein flow cytometry
- Flow cytometry lecture
- Terahertz spectroscopy principles and applications
- Atomic emission spectroscopy principle
- Iccs 2016 cytometry
- Limitations of the beer lambert law
- Rfu fluorescence
- Fluorescence microscopy uses
- Jablonski
- Chemiluminescence vs fluorescence
- Confocal fluorescence microscopy
- Bright field microscopy
- Cold vapor atomic fluorescence spectrometry
- Light sources for fluorescence microscopy
- Cold vapor atomic fluorescence spectrometry
- Fluorescence activated cell sorting
- Protein fragment complementation assay
- Fluorescence bandpass filter
- Biotek flx800 fluorescence microplate reader
- Jablonski diagram fluorescence
- Fluorescence recovery after photobleaching
- Fluorescence microscopy
- Fluorescence recovery after photobleaching
- Jablonski diagram
- Fluorescence-activated cell sorting (facs)
- Magic box respiratory
- Simple mask flow rate
- High flow versus low flow oxygen