THEMIS INSTRUMENT SUITE FM 4FM 5 PRESHIP REVIEW
- Slides: 21
THEMIS INSTRUMENT SUITE FM 4/FM 5 PRE-SHIP REVIEW INTRODUCTION Davin Larson SST Lead University of California - Berkeley FM 4/FM 5 IPSR SST 1 UCB, May 5 th, 2006
Outline • Summary of SST Status at FM 1 PSR • FM 1 -5 Configuration by SN • End-Item Data Package • Final Test and Cal Results vs. Requirement Fulfillment • Open Items FM 4/FM 5 IPSR SST 2 UCB, May 5 th, 2006
Status Current SST-DAP Status: FM 1 - Delivered FM 2 - Delivered FM 3 - Delivered FM 4 - Delivered FM 5 - Delivered FM 6 - Delivered (Needs memory test following conformal coating) FM 4/FM 5 IPSR SST 3 UCB, May 5 th, 2006
Status Current SST-Sensor Status: FM 1 – (S 01 & S 02) Delivered FM 2 - (S 06 & S 04) Delivered FM 3 - (S 05 & S 10) Delivered FM 4 - (S 07 & S 09) Delivered FM 5 - (S 11 & S 12) Complete, Calibrated, awaiting delivery. FM 6 - (S 03 & S 08) Refurbished, requires workmanship vibe. FM 4/FM 5 IPSR SST 4 UCB, May 5 th, 2006
Flight Configuration • Current flight configuration of sensors and DAPs for all flight units. FM 4/FM 5 IPSR SST 5 UCB, May 5 th, 2006
FM 1 EIDP: Sensors • THEMIS SST Sensor: (Sensor 1: Binder #33; Sensor 2: Binder #34) • Electrical Assembly • Schematics • Electrical Assembly Records • Rework/Repair history • Electrical test results • QA signoffs • Mechanical Assembly • Mechanical drawings • Vibration test results • Thermal Vac test results • Assembly Records. • Special TVAC Acceptance Test Data. FM 4/FM 5 IPSR SST 6 UCB, May 5 th, 2006
FM 1 EIDP: DAP • THEMIS SST DAP: (Binder#2) • Schematics • Electrical Assembly Records • Rework/Repair history • Electrical test results • QA signoffs FM 4/FM 5 IPSR SST 7 UCB, May 5 th, 2006
SST Vibration Testing (Pictures of Probe 2 units) FM 4/FM 5 IPSR SST 8 UCB, May 5 th, 2006
SST Acceptance Level Dynamic Test SST Serial Number Test Date Following Initial Build Retest Date (Rework/Redesign) 13 Jan 05 Oct 11, 2005 (full-qual) 22 Apr 05 Nov 4, 2005 (workmanship) June 13, 2005 Nov 4, 2005 (workmanship) May 6 2005 Nov 7, 2005 (workmanship) Nov 7 2005 Not needed Dec 13 2005 Not needed 001, 002 003, 004 006, 008 007, 009 011, 012 005, 010 FM 4/FM 5 IPSR SST 9 UCB, May 5 th, 2006
SST Acceptance Level Dynamic Test Per UCB THEMIS Document THM-SST-PRC-002 • • • All 3 Axes Sine Survey, 0. 5 G, 5 – 2000 Hz Sine Strength per Swales TM 2430 -Rev. D, 29 November 2004 Random – Per Swales Document TM 2430 -Rev. D – 12. 51 G RMS – 1 min duration per axis All Vibration Testing Performed at Quanta Laboratories Retesting (if required) • • ‘Workmanship Shake’ – Validate any design changes following original Flight build shake – Validate rework of any failures on previous vibration test Entire vibration protocol, single axis only Acceptance Sine Strength Specification FM 4/FM 5 IPSR SST Flight Unit SST 10 Acceptance Level Random Specification UCB, May 5 th, 2006
SST TVAC Testing SST Serial Number Test Date Following Initial Build Retest Date (Rework/Redesign) 5 Jan 05 Not needed 28 Apr 05 Not needed Sept 7, 2005 Dec-3 -2005 (sensor 6 only) May 25, 2005 not needed Dec 3 2005 not needed Dec 20, 2005 to Jan 11, 2006 (8 cycles) not needed 001, 002 003, 004 006, 008 007, 009 011, 012 005, 010 FM 4/FM 5 IPSR SST 11 UCB, May 5 th, 2006
Thermal Profile Procedure: THM-SST-PRC-003 FM 4/FM 5 IPSR SST 12 UCB, May 5 th, 2006
SST -Magnetic cleanliness SST Magnetic Cleanliness Testing Status • All units have been tested and are nominal. FM 4/FM 5 IPSR SST 13 UCB, May 5 th, 2006
SST -Magnetic cleanliness SST Sensor 07 • All units have been tested and are nominal. FM 4/FM 5 IPSR SST 14 UCB, May 5 th, 2006
SST -Magnetic cleanliness SST Sensor 09 FM 4/FM 5 IPSR SST 15 UCB, May 5 th, 2006
SST -Magnetic cleanliness SST Sensor 11 FM 4/FM 5 IPSR SST 16 UCB, May 5 th, 2006
SST -Magnetic cleanliness SST Sensor 12 FM 4/FM 5 IPSR SST 17 UCB, May 5 th, 2006
SST Requirements SST Mission Requirements Fulfillment: • FM 1 MR fulfillment noted in THEMIS Verification Matrix (see MSE Presentation) • Integrated SST and component parts compliant with mission requirements FM 4/FM 5 IPSR SST 18 UCB, May 5 th, 2006
Open PFRs None FM 4/FM 5 IPSR SST 19 UCB, May 5 th, 2006
Recent PFRs PFR 169 • Found poor ion energy response in 2 sensors. (S 03 and S 08). Found only in sensors that had undergone electron testing. • Discovered that plasma discharges from the electron gun had inadvertently damaged adjacent ion detectors. (electron detectors where protected by foil that stopped all ions and never showed any damage. ) • Subsequent testing with the electron gun is done with the open (ion) detectors covered. Electron gun has also been refurbished. • Swapped affected sensors with spares (S 05 and S 10). • Replaced damaged open (ion) detectors in affected units. These are now the new spares. PFR is closed. FM 4/FM 5 IPSR SST 20 UCB, May 5 th, 2006
Results of Testing/Calibration • • Energy Threshold vs temperature measured and acceptable Energy resolution acceptable (<13 ke. V at 5 C). Detector gain measured and is stable over temperature. Channel crosstalk measured and acceptable (<20 ke. V) Attenuator functions over full operational Temp range. Heater/Thermostat Functional at < -50 C *! Magnetic moments measured and deemed acceptable. End to End performance verified. • (*!) Noise susceptibility from BAU is very low now FM 4/FM 5 IPSR SST 21 UCB, May 5 th, 2006
- Preship
- Themis fm
- Spedas wiki
- Themis constellation
- Themis athanassiadou
- Esa themis
- Themis
- Katrina connors
- Themis
- Themis
- Themis software
- Themis
- Themis pi
- Themis athanassiadou
- Themis
- Themis dc
- Themis
- Themis lenguaje inclusivo
- Themis competition
- Active vs passive instruments
- Qualitative assessment and review instrument
- Traditional and systematic review venn diagram