FNAL Hybrid Thermal Test Status Hybrid Thermal Test
FNAL Hybrid Thermal Test Status ØHybrid Thermal Test Procedure ØTest Results Eva Medel FNAL CMS Group June 22, 2004 May 13, 2004
Hybrid Thermal Test Procedure We follow exactly the 4 Hybrid Thermal Test Procedure; basically: I. Pre-Loading Turn the chiller on, activate ACDC, NIM crate and peltier power supply power on. II. Loading Load the hybrids onto the peltier plate. Hybrid registration. Turn on ARC Board power supplies. Initialization in testation_start. vi. III. Testing (failures) üOne or more shorted channels, or üFive or more noisy or open channels. May 13, 2004
Hybrid Thermal Test Procedure After the test, an automatic report is generated for each hybrid in teststation_final_report. vi xml files generated and saved in the same directories, ( e. g. C: Traker. DBresults). Uploading of the xml files into the data base when the program arcs_auto. xml. exe is run. May 13, 2004
Hybrid Thermal Test Results 430 TEC hybrids have been tested ü 69 TEC hybrids failed thermal test ( 16% ) ü 361 passed the test ( 84% ) Hybrids that failed ü 44 showed at – 20 PLL failure. ü 25 for other reasons May 13, 2004
Hybrid Thermal Test Results 210 TOB hybrids have been tested ü 29 TOB hybrids failed thermal test ( 14% ) ü 181 passed the test ( 86% ) Hybrids that failed ü 24 showed at – 20 PLL failure. ü 5 for other reasons May 13, 2004
Hybrid Thermal Test TEC HYBRIDS Failure at – 200 C. 22 do not present any apparent damage. The rest show certain damage mainly done during the wire-bonding like: ü OUTER BONDS REPAIRS ü NEW WIRE, NEW TOOL, PROGRAM MODE-SOFTWARE UPGRADES ü IRREGULAR LOOP HEIGHTS Or problems due to PA dirty, scratched or bad chips. May 13, 2004
Hybrid Thermal Test TEC HYBRIDS For failed hybrids ( except – 20 PLL failure) we used Lance’ s Labview program to know how noisy is an APV, ploting the noise and pedestal vs channel. v This tool allow us to determine short channels. The cut applied on the calibration signal height was 1. 7 ADC counts. Short channels due to the wire bonding. v To determine open channels we analyze the pedestal distribution expecting a smooth spectrum. Open channels are attributed to open bonds and PA damages. A visual inspection, clean some hybrids and retest were done. May 13, 2004
Hybrid Thermal Test TEC HYBRIDS Example of short and noisy channels May 13, 2004
Hybrid Thermal Test TEC HYBRIDS May 13, 2004
Hybrid Thermal Test TEC HYBRIDS Results ü Using Lance’ s program we found less short channels than reported. ü Noisy channels are consistent with the reported ones. ü Visual inspection was done and then some hybrids were retested. The majority failed for the same reason than the first time, ( no visible problem). ü 83% of the failed are 6 chip hybrids. The number of short channels should not be the same for 4 and 6 chip hybrids. May 13, 2004
Hybrid Thermal Test TOB HYBRIDS Ø 29 TOB hybrids that failed thermal test Ø 25 showed at – 20 PLL failure Ø For the rest of the TOB hybrids (4) an analysis using Lance’ s program was done, and visual inspection trying to fix something. May 13, 2004
Hybrid Thermal Test TOB HYBRIDS Failure at – 200 C. 10 do not present any apparent damage. The rest show certain damage on the pitch adapter mainly: ü SCRATCHES ON PA ü BAD SURFACE (bond pads, 1) May 13, 2004
Hybrid Thermal Test TOB HYBRIDS A hybrid without problems (peddecon_file 1) May 13, 2004
Hybrid Thermal Test TOB HYBRIDS A hybrid without problems (peddecon_file 2) May 13, 2004
Hybrid Thermal Test TOB HYBRIDS A hybrid without problems (peddecon_file 3) May 13, 2004
Hybrid Thermal Test TOB HYBRIDS May 13, 2004