Thermal Compensation System for Virgo Alessio Rocchi INFN

  • Slides: 21
Download presentation
Thermal Compensation System for Virgo+ Alessio Rocchi INFN Roma Tor Vergata A. Rocchi -

Thermal Compensation System for Virgo+ Alessio Rocchi INFN Roma Tor Vergata A. Rocchi - 24. 10. 2007 LSC-Virgo Meeting

Thermal Lensing • Observed in Virgo through reduction of the sidebands gain. A. Rocchi

Thermal Lensing • Observed in Virgo through reduction of the sidebands gain. A. Rocchi - 24. 10. 2007 LSC-Virgo Meeting

ITMs absorptions evaluated through coupled thermal-modal FEM with Ansys Red and blue curves experimental

ITMs absorptions evaluated through coupled thermal-modal FEM with Ansys Red and blue curves experimental frequency shifts green and pink FEM results 0. 025 Df (Hz) WI NI New measurement by M. Punturo, NI absorption ~ 4. 5 ppm 0 Time span ~25000 sec ITMs effective absorptions WI (7. 8± 1. 1) ppm NI (2. 3± 0. 3) ppm Errors come from dependency of the model from uncertainty on elastic parameters dependency on temperature A. Rocchi - 24. 10. 2007 LSC-Virgo Meeting

Thermal lensing evaluation Beausoleil analitycal solution A. Rocchi - 24. 10. 2007 LSC-Virgo Meeting

Thermal lensing evaluation Beausoleil analitycal solution A. Rocchi - 24. 10. 2007 LSC-Virgo Meeting

Thermal Compensation System • • TCS for Virgo+: annular heating of ITMs with dc

Thermal Compensation System • • TCS for Virgo+: annular heating of ITMs with dc CO 2 laser “Annular” profile – from simulations: reasonable parameters to compensate inner radius~2. 5 cm, outer radius~11÷ 14 cm Desirable properties: • It is easily adaptable as new understanding of the ITF is realized • It does not require a significant vacuum incursion to install as this would lead to significant down time to the instrument Two phases installation: 1. Feb 2008, system without power stabilization control loop 2. May-June 2008, installation of the power stabilization control loop A. Rocchi - 24. 10. 2007 LSC-Virgo Meeting

TCS Power requirements Virgo on WI mirror (absorption 7. 7 ppm) ~1. 8 W

TCS Power requirements Virgo on WI mirror (absorption 7. 7 ppm) ~1. 8 W Virgo (nominal absorption) ~0. 4 W Virgo on WI mirror (absorption 7. 7 ppm, Plas=25 W) ~6. 2 W Virgo (nominal absorption, Plas=25 W) ~1. 3 W Virgo+, (F=150, Grec=19) (nominal absorption) ~1. 5 W Virgo+ (F=150, Grec=35) (nominal absorption) ~2. 8 W A. Rocchi - 24. 10. 2007 LSC-Virgo Meeting

Optical layout A. Rocchi - 24. 10. 2007 LSC-Virgo Meeting

Optical layout A. Rocchi - 24. 10. 2007 LSC-Virgo Meeting

Optical benches positioning NI tower Acoustic enclosure TCS optical bench WI tower A. Rocchi

Optical benches positioning NI tower Acoustic enclosure TCS optical bench WI tower A. Rocchi - 24. 10. 2007 LSC-Virgo Meeting Oven

Final design of the Optical Imaging System Single axicon to get the “annular” profile

Final design of the Optical Imaging System Single axicon to get the “annular” profile P. La Penna & E. Genin 250 mm 100% efficiency A. Rocchi - 24. 10. 2007 LSC-Virgo Meeting

UHV Zn. Se viewport DN 100 CF Clear aperture 90 mm Two large in-vacuum

UHV Zn. Se viewport DN 100 CF Clear aperture 90 mm Two large in-vacuum steering mirrors (f=190 mm) Positioning controlled manually in all DOFs Remotely controlled via stepping motors in angular DOFs and one translation A. Rocchi - 24. 10. 2007 LSC-Virgo Meeting

Schematic view Beam Dump CO 2 Laser output power control l/2 waveplate Polarizer L

Schematic view Beam Dump CO 2 Laser output power control l/2 waveplate Polarizer L 1 L 2 AOM Beam Dump Power Meter L 3 AXICON Components in the blue box are used for laser power stabilization, will be installed by mid 2008 L 4 L 5 LX Aiming Laser A. Rocchi - 24. 10. 2007 LSC-Virgo Meeting To in-vacumm steering mirror

Tests in Tor Vergata Labs before installation on Virgo • Measure the RIN of

Tests in Tor Vergata Labs before installation on Virgo • Measure the RIN of the laser – So we can have an estimate of the injected noise in the first phase installation • Check the efficiency of the Optical Imaging System A. Rocchi - 24. 10. 2007 LSC-Virgo Meeting

Preliminary design of the first in-vacuum mirror ~800 mm Material: Aluminum T. Zelenova A.

Preliminary design of the first in-vacuum mirror ~800 mm Material: Aluminum T. Zelenova A. Rocchi - 24. 10. 2007 LSC-Virgo Meeting

Exploring different designs for the in-vacuum optics holders, with T. Zelenova 226 Hz 252

Exploring different designs for the in-vacuum optics holders, with T. Zelenova 226 Hz 252 Hz 537 Hz A. Rocchi - 24. 10. 2007 LSC-Virgo Meeting

f 0=537 Hz Unloaded f 0=375 Hz Loaded with mirror (2 kg) A. Rocchi

f 0=537 Hz Unloaded f 0=375 Hz Loaded with mirror (2 kg) A. Rocchi - 24. 10. 2007 LSC-Virgo Meeting

Error signal generation • • • E. Calloni & M. Laval demonstrated that Phase

Error signal generation • • • E. Calloni & M. Laval demonstrated that Phase camera 0 + Dark. F tools to “measure” TCS Pine-hole photodiodes error signal generation Phase camera 1 an upgraded tool to understand sidebands and TCS B 5 Pin-hole mirror Photodiodes P 1 Signal = P 1 -P 2 This is only one error signal, but two are needed. Investigating the possibility to use the signal from B 2 A. Rocchi - 24. 10. 2007 LSC-Virgo Meeting

CO 2 laser power stabilization 20 W CO 2 Laser New LASY-20 S from

CO 2 laser power stabilization 20 W CO 2 Laser New LASY-20 S from Access Laser Co, pre-stabilized, temperature tuned (no line hops) PD AOM Critical components PD A. Rocchi - 24. 10. 2007 LSC-Virgo Meeting

Noise budget Evaluation of TCS noise injected in Virgo+, if RIN is stabilized @

Noise budget Evaluation of TCS noise injected in Virgo+, if RIN is stabilized @ 3× 10 -7/√Hz, flat over the frequency range 10 Hz to 1000 Hz, TCS power=1. 5 W. Virgo+ sensitivity (without monolithic suspensions) Virgo+ sensitivity (with monolithic suspensions) Total Flexural Thermoelastic+Thermorefractive Radiation pressure A. Rocchi - 24. 10. 2007 LSC-Virgo Meeting

A. Rocchi - 24. 10. 2007 LSC-Virgo Meeting

A. Rocchi - 24. 10. 2007 LSC-Virgo Meeting

Building and validating the FE model Comparison with analytical solution Beausoleil (LIGO-P 020026 -00

Building and validating the FE model Comparison with analytical solution Beausoleil (LIGO-P 020026 -00 -D) Max difference ~ 0. 3% • • • DRUM MODE ANSYS FEM f= 5584. 15 Hz Experimental value f= 5584. 7 Hz Difference is 0. 55 Hz less than 0. 01% A. Rocchi - 24. 10. 2007 LSC-Virgo Meeting

Possible optical layout AXICON A. Rocchi - 24. 10. 2007 LSC-Virgo Meeting

Possible optical layout AXICON A. Rocchi - 24. 10. 2007 LSC-Virgo Meeting