YL 6500 GC THE i DEA MAKES i

  • Slides: 20
Download presentation
YL 6500 GC THE i. DEA MAKES i. DEAL

YL 6500 GC THE i. DEA MAKES i. DEAL

The History of GC Development 680 A M 600 D 1991 Acme 6000 2011

The History of GC Development 680 A M 600 D 1991 Acme 6000 2011 INNOVATIVE YL 6500 GC 1995 2001 YL 6100 2006

YL 6500 GC Features YL 6500 GC comes with Enhanced 4 Vs • View

YL 6500 GC Features YL 6500 GC comes with Enhanced 4 Vs • View • Verification • Variety • Value

YL 6500 GC Features Enjoy the View • Enlarged Haptic color LCD(5. 7”) showing

YL 6500 GC Features Enjoy the View • Enlarged Haptic color LCD(5. 7”) showing all factors at a glance. • Classic and lab-suitable Design • Intuitive and user-friendly view by YL chromatography data system

YL 6500 GC Features Reliable Verification • Strongly Enhanced APCs for almost all detectors

YL 6500 GC Features Reliable Verification • Strongly Enhanced APCs for almost all detectors - Increased precision in pressure and flow rate - Automatic compensation for temperature & pressure in installation condition - Pressure setpoint precision to 0. 01 psi - Shockproof APCs with short flow length - 4 column flow control modes: constant pressure, constant flow, programmed pressure (5 steps), programmed flow (5 steps). - Stable APCs against oven temperature changes - Up to six APCs can be installed and up to 18 channels of APC.

YL 6500 GC Features Reliable Verification • High speed data process by network (LAN)

YL 6500 GC Features Reliable Verification • High speed data process by network (LAN) communication • Ten heated Zones to enhance a precision : 1 heated oven zone/ 3 heated injector zone 3 heated detector zone/ 2 heated valve zone 1 heated methanizer zone • Powerful and easy control by YL chromatography software

YL 6500 GC Features Wide Variety • Maximum no. of inlet / detector to

YL 6500 GC Features Wide Variety • Maximum no. of inlet / detector to be installed : 3 • Various inlets available - Capillary Inlet - Packed Inlet - On-Column Inlet (Temperature programming up to 5 steps)

YL 6500 GC Features Wide Variety • Various detectors with high sensitivity available -

YL 6500 GC Features Wide Variety • Various detectors with high sensitivity available - Flame Ionization Detector (FID) : MDL < 2. 0 pg C/sec - Thermal Conductivity Detector (TCD) : MDL < 2. 5 ng /ml - Micro- Thermal Conductivity Detector (μTCD) : MDL < 400 pg/ml - Nitrogen Phosphorus Detector(NPD) - Flame Photometric Detector (FPD) - Electron Capture Detector(ECD) : MDL < 10 fg/sec - Pulsed discharge Detector (PDD) - Pursed Frame Photometric Detector (PFPD)

YL 6500 GC Features Upgrade the Value • Reasonable price saving user’s valuable time

YL 6500 GC Features Upgrade the Value • Reasonable price saving user’s valuable time • Improved column conditioning function : Automatic set of split flow up to 5 ml/min on column conditioning • Sensitive leak detection (H 2) • Prevention of oven malfunction (Over heating) • Automatic stop when an oven door gets opened during operation. • Time control: Automatic turn-on/off

YL 6500 GC Modules Superior Performance in Column Oven • Usable volume : 14

YL 6500 GC Modules Superior Performance in Column Oven • Usable volume : 14 L • Temperature operating range : (ambient to 450℃ or more) : - 80℃ ~ 450℃ (with LN 2 cryogenic cooling) : - 55℃ ~ 450℃ (with CO 2 cryogenic cooling) • Temperature set-point : 1℃ • Programming: 25 ramps, 26 steps temperature program • Maximum set-point heating rate: 100℃/min • Cool down rate: 6. 5 minutes from 450℃ to 50℃ • Temperature program method: Maximum up to 20 • Temperature stability : ± 0. 01℃ (Isothermal)

YL 6500 GC Modules Enhanced Performance in Inlets • Packed inlet - Maximum Temperature

YL 6500 GC Modules Enhanced Performance in Inlets • Packed inlet - Maximum Temperature : 450 o. C - Pressure setting range : 0. 01 ~ 100 psi - Total flow setting range : 0. 1~100 ml/min - Flow stability < ± 0. 05 ml/min - Pressure stability < ± 0. 05 psi - Temperature setpoint : 1℃ - Temperature stability < ± 0. 1℃

YL 6500 GC Modules Enhanced Performance in Inlets • Capillary inlet (Split/Splitless) - Maximum

YL 6500 GC Modules Enhanced Performance in Inlets • Capillary inlet (Split/Splitless) - Maximum temperature : 400 o. C - Pressure setting range : 0. 01 ~ 100 psi - Total flow setting range : 0. 1 ~ 400 ml/min N 2 0 ~ 1000 ml/min He - Splitless time setpoint : 0. 01 min - Total flow stability < ± 0. 05 ml/min - Pressure stability < ± 0. 05 psi - Temperature setpoint : 1℃ - Temperature stability < ± 0. 1℃

YL 6500 GC Modules Enhanced Performance in Inlets • On-column Inlet - Maximum Temperature

YL 6500 GC Modules Enhanced Performance in Inlets • On-column Inlet - Maximum Temperature : 450 o. C - Pressure setting range : 0. 01 ~ 100 psi - Total flow setting range : 0. 1 ~ 100 ml/min - Flow stability < ± 0. 05 ml/min - Pressure stability < ± 0. 05 psi - Temperature setpoint : 1℃ - Temperature stability < ± 0. 1℃ - Temperature programming up to 5 steps

YL 6500 GC Modules Exceptional Sensitivity in Detectors • Flame Ionization Detector - Maximum

YL 6500 GC Modules Exceptional Sensitivity in Detectors • Flame Ionization Detector - Maximum operating temperature : 450 ℃ - Automatic flame ignition - Temperature set-point : 1 ℃ - Ignition message: Alarm message as ‘Flame out’ - MDL : 2. 0 pg carbon/sec - Linear dynamic range : 107 -Temperature stability : ± 0. 1℃ - Data acquisition rate : 200 Hz

YL 6500 GC Modules Exceptional Sensitivity in Detectors • Thermal Conductivity Detector - Maximum

YL 6500 GC Modules Exceptional Sensitivity in Detectors • Thermal Conductivity Detector - Maximum operating temperature : 400 ℃ - Flow through cell : 4 Rhenium-Tungsten filaments - Temperature set-point : 1 ℃ - Filament protection - MDL : 2. 5 ng/ml - Micro-cell (Option) : μTCD (MDL < 400 pg/ml) - Data acquisition rate : 200 Hz

YL 6500 GC Modules Exceptional Sensitivity in Detectors • Nitrogen Phosphorous Detector - Maximum

YL 6500 GC Modules Exceptional Sensitivity in Detectors • Nitrogen Phosphorous Detector - Maximum operating temperature : 400 ℃ - MDL : <0. 4 pg N/sec (Azobenzene) <0. 2 pg P/sec (Malathion) - Dynamic Range for N : > 104 - Dynamic Range for P : > 104

YL 6500 GC Modules Exceptional Sensitivity in Detectors • Flame Photometric Detector - Maximum

YL 6500 GC Modules Exceptional Sensitivity in Detectors • Flame Photometric Detector - Maximum operating temperature : 300 ℃ - Minimum detection limit for S : < 20 pg S/sec - Minimum detection limit for P : < 0. 5 pg P/sec - Dynamic Range for S : > 103 - Dynamic Range for P : > 104 - S/C Selectivity : 105 - P/C Selectivity : 106

YL 6500 GC Modules Exceptional Sensitivity in Detectors • Pulsed Discharge Detector - Maximum

YL 6500 GC Modules Exceptional Sensitivity in Detectors • Pulsed Discharge Detector - Maximum operating temperature : 400 ℃ - PDHID, PDECD mode is selectable - Helium ionization mode(PDHID) : Linearity – 105 MDL (organic compound : low ppb) (permanent gas : low ppm) - Electron capture mode(PDECD) : Linearity – 105 MDL – 10 -15 Dopant gas : 3% Xe in Helium

YL 6500 GC Competition • Specifications of YL 6500 GC VS “A” company YL

YL 6500 GC Competition • Specifications of YL 6500 GC VS “A” company YL 6500 GC “A” 7890 A Maximum Number of Heated Zones 10 7 Maximum Number of Inlet Maximum Number of Detectors Maximum GC Methods Storage Maximum Programming Ramp/Plateaus 3 3 20 2 3 5 25/26 20/21 Oven Cooldown Rate 6. 5 min 450℃ to 50℃ 4. 0 min 450℃ to 50℃ FID MDL TCD MDL ECD MDL 2. 0 pg carbon/sec as 1. 8 pg C/s (for tridecane) dodecane using N 2 carrier 2. 5 ng/ml as dodecane 0. 4 ng tridecane/m. L μTCD : 0. 4 ng/ml with He carrier <10 fg/sec as Lindane <6 fg/m. L Lindane

YL 6500 GC Competition • Specifications of YL 6500 GC VS “P” company YL

YL 6500 GC Competition • Specifications of YL 6500 GC VS “P” company YL 6500 GC “A” Clarus 680 Maximum Number of Heated Zones 10 Maximum Number of Inlet Maximum Number of Detectors Maximum GC Methods Storage Maximum Programming Ramp/Plateaus 3 3 20 2 2 5 25/26 9/10 Oven Cooldown Rate 6. 5 min 450℃ to 50℃ 2. 0 min 450℃ to 50℃ in a specific condition FID MDL 2. 0 pg carbon/sec as dodecane using N 2 carrier 2. 5 ng/ml as dodecane μTCD : 0. 4 ng/ml TCD MDL ECD MDL <10 fg/sec as Lindane 3 pg C/s (for tridecane) 1000 ng/m. L nonane < 0. 05 pg nitrogen