THE DICOM 2013 INTERNATIONAL CONFERENCE SEMINAR March 14

  • Slides: 14
Download presentation
THE DICOM 2013 INTERNATIONAL CONFERENCE & SEMINAR March 14 -16 Bangalore, India On handling

THE DICOM 2013 INTERNATIONAL CONFERENCE & SEMINAR March 14 -16 Bangalore, India On handling low-dose CT images in the absence of reliable DICOM header information Laks Raghupathi Ph. D. Veerendra Shetty Presenter: Laks Raghupathi Siemens Corporate Development Center, Bangalore, India Healthcare 1 – System House, Algorithms Research Group Matthias Wolf Ph. D. Siemens Medical Solutions USA, Inc. , Malvern, PA, USA SYNGO R&D US Innovation March 2013 DICOM International Conference & Seminar Copyright © Siemens AG 2013. All rights reserved.

Handling Low Dose CT • Background • Capturing Dosage in DICOM • Proposed Approach

Handling Low Dose CT • Background • Capturing Dosage in DICOM • Proposed Approach & Methodology • Results & Discussion • Author Correspondence March 2013 DICOM International Conference & Seminar L. Raghupathi et al. , On handling low-dose CT images in the absence of reliable DICOM header information Copyright © Siemens AG 2013. All rights reserved. 2

Background • CT is important diagnostic imaging modality with millions of procedures performed each

Background • CT is important diagnostic imaging modality with millions of procedures performed each year • Modern image acquisition technologies are intended to enable dose reduction in CT imaging while maintaining diagnostic information • Evolving technology with newer scanners and reconstruction kernels. • Low dose CT (LDCT) tend to have higher noise and less sharpness • LDCT may affect performance of advanced post-processing algorithms, such as Computer-aided Detection (CAD) March 2013 DICOM International Conference & Seminar L. Raghupathi et al. , On handling low-dose CT images in the absence of reliable DICOM header information Copyright © Siemens AG 2013. All rights reserved. 3

Capturing Dosage in DICOM • The first step is to know the dosage for

Capturing Dosage in DICOM • The first step is to know the dosage for each imaging procedure • “Exposure”, “XRay. Tube. Current”, “KV” capture input attributes • CTDose. Index field captures overall dosage • Non-Type 1 attributes – not mandatory • Proposal: Radiation Dose Structured Report (RDSR) • CTDIvol (m. Gy), DLP (m. Gy * cm), etc. captures dosage • All new devices to encode dose in RDSR [1] • Drawbacks: • Old scanners may not be updated with RDSR • Legacy PACS systems may not read this new SR object __ [1] D. Clunie, Extracting, Managing and Rendering DICOM Radiation Dose Information from Legacy & Contemporary CT Modalities, DICOM Intl. Conf. , 2010 March 2013 DICOM International Conference & Seminar L. Raghupathi et al. , On handling low-dose CT images in the absence of reliable DICOM header information Copyright © Siemens AG 2013. All rights reserved. 4

Proposed Approach • Smoothing reduces noise, but in excess removes information à Algorithms like

Proposed Approach • Smoothing reduces noise, but in excess removes information à Algorithms like Lung CAD* expect sharp kernel Soft kernels may affect performance such as high false positive (FP) rates [2] • Need to strike balance between noise and sharpness • New Idea: Intelligent pre-processing filter to adjust image quality __ [2] M. Yanagawa et al. , Pulmonary nodules: Effect of ASIR technique on performance of a CAD system-comparison of performance between different-dose CT scans Eur. J. Rad. , 81(10) 2012 __ * Lung CAD research prototype is not commercially available for sale in the USA or elsewhere and is not related to any product. March 2013 DICOM International Conference & Seminar L. Raghupathi et al. , On handling low-dose CT images in the absence of reliable DICOM header information Copyright © Siemens AG 2013. All rights reserved. 5

Framework Scanner 1 Scanner 2 Intelligent Pre. Processing* [3] Post. Processing Applications (e. g.

Framework Scanner 1 Scanner 2 Intelligent Pre. Processing* [3] Post. Processing Applications (e. g. , CAD) Reading Workplace PACS __ [3] L. Raghupathi et al. , An Intelligent Pre-Processing Framework for Standardizing Medical Images for CAD and Other Post. Processing Applications, Proc. SPIE Med. Imag. , 2012 __ *Research prototype is not commercially available for sale in the USA or elsewhere and is not related to any product. March 2013 DICOM International Conference & Seminar L. Raghupathi et al. , On handling low-dose CT images in the absence of reliable DICOM header information Copyright © Siemens AG 2013. All rights reserved. 6

Choosing Filters Original Filter 1 Filter 2 Filter 3 Filter 4 It is possible

Choosing Filters Original Filter 1 Filter 2 Filter 3 Filter 4 It is possible to adjust sharpness using the filters. Key question: Which filter is best for Lung CAD* performance? __ * Lung CAD research prototype is not commercially available for sale in the USA or elsewhere and is not related to any product. March 2013 DICOM International Conference & Seminar L. Raghupathi et al. , On handling low-dose CT images in the absence of reliable DICOM header information Copyright © Siemens AG 2013. All rights reserved. 7

Methodology Signal/Contrast to Noise Ratio dependent on location Use image metrics for “adaptive” pre-processing

Methodology Signal/Contrast to Noise Ratio dependent on location Use image metrics for “adaptive” pre-processing Estimate noise (N) around air region (carina landmark) Estimate sharpness (S) in air-body border Iteratively filter till desired Noise to Sharpness ratio (NSR) reached March 2013 DICOM International Conference & Seminar L. Raghupathi et al. , On handling low-dose CT images in the absence of reliable DICOM header information Copyright © Siemens AG 2013. All rights reserved. 8

March 2013 DICOM International Conference & Seminar Fi lte r 5 Fi lte r

March 2013 DICOM International Conference & Seminar Fi lte r 5 Fi lte r 4 Fi lte r 3 Fi lte r 2 Fi lte r 1 O rig in al NSR Outputs L. Raghupathi et al. , On handling low-dose CT images in the absence of reliable DICOM header information Copyright © Siemens AG 2013. All rights reserved. 9

LDCT Details & Ground Truth • 34 Siemens IRIS Kernel avg. 90 m. As,

LDCT Details & Ground Truth • 34 Siemens IRIS Kernel avg. 90 m. As, 40 solid pulmonary nodules (SPN) in 4 -30 mm dia. • 88 Siemens IRIS Kernel simulated at 50, 100 m. As, 67 SPN in 4 -30 mm dia. March 2013 DICOM International Conference & Seminar L. Raghupathi et al. , On handling low-dose CT images in the absence of reliable DICOM header information Copyright © Siemens AG 2013. All rights reserved. 10

Results Trained on large-scale CT datasets Validation on low-dose CT images at different levels

Results Trained on large-scale CT datasets Validation on low-dose CT images at different levels Improved performance* (mean FP down 66%, median down 50%) __ * Lung CAD research prototype is not commercially available for sale in the USA or elsewhere and is not related to any product. March 2013 DICOM International Conference & Seminar L. Raghupathi et al. , On handling low-dose CT images in the absence of reliable DICOM header information Copyright © Siemens AG 2013. All rights reserved. 11

Discussion • Proposed approach normalizes images with different dosage • Uses measured attributes independent

Discussion • Proposed approach normalizes images with different dosage • Uses measured attributes independent of acquisition parameters • Demonstrated potential for improved CAD performance on LDCT • Possible to extend to other post-processing applications • Potential to enhance their applicability in clinical workflows • Practical alternative till DICOM dosage attributes captured March 2013 DICOM International Conference & Seminar L. Raghupathi et al. , On handling low-dose CT images in the absence of reliable DICOM header information Copyright © Siemens AG 2013. All rights reserved. 12

Acknowledgements • Marc Rose – Healthcare 1, Siemens CT DC • Dr. Sean Zhou

Acknowledgements • Marc Rose – Healthcare 1, Siemens CT DC • Dr. Sean Zhou and Ingo Schmuecking – SYNGO R&D US March 2013 DICOM International Conference & Seminar L. Raghupathi et al. , On handling low-dose CT images in the absence of reliable DICOM header information Copyright © Siemens AG 2013. All rights reserved. 13

Correspondence Veerendra Shetty • • Veerendra. Shetty@siemens. com Head System Function House – Healthcare

Correspondence Veerendra Shetty • • Veerendra. Shetty@siemens. com Head System Function House – Healthcare 1 Siemens Technology & Services 84, Keonics Electronics City, Bangalore, India Thank you for your attention ! March 2013 DICOM International Conference & Seminar L. Raghupathi et al. , On handling low-dose CT images in the absence of reliable DICOM header information Copyright © Siemens AG 2013. All rights reserved. 14