TAPPINGMODE IMAGING APPLICATIONS AND TECHNOLOGY Tapping Mode imaging

  • Slides: 30
Download presentation
TAPPINGMODE™ IMAGING APPLICATIONS AND TECHNOLOGY Tapping. Mode imaging is implemented in ambient air (or

TAPPINGMODE™ IMAGING APPLICATIONS AND TECHNOLOGY Tapping. Mode imaging is implemented in ambient air (or liquid) by oscillating the cantilever assembly at or near the cantilever’s resonant frequency using a piezoelectric crystal. The piezo motion causes the cantilever to oscillate with a high amplitude (the “free air” amplitude, typically greater than 20 nm) when the tip is not in contact with the surface. The oscillating tip is then moved toward the surface until it begins to lightly touch, or “tap” the surface. During scanning, the vertically oscillating tip alternately contacts the surface and lifts off, generally at a frequency of 50, 000 to 500, 000 cycles per second. As the oscillating cantilever begins to intermittently contact the surface, the cantilever oscillation is necessarily reduced due to energy loss caused by the tip contacting the surface. The reduction in oscillation amplitude is used to identify and measure surface features.

TM AFM: Constant Force Gradient: taking advantage of the vibrational characteristics of the cantilever

TM AFM: Constant Force Gradient: taking advantage of the vibrational characteristics of the cantilever

ACantilever = Acantilever sin( t+ 1) Resonant Applied electronic Signal to the Piezo V

ACantilever = Acantilever sin( t+ 1) Resonant Applied electronic Signal to the Piezo V piezo = Vac sin( t+ 1) Produced a sine wave oscillation with the same frequency and phase angle A piezo = Aac sin( t+ 1)

Tapping cantilever on free air

Tapping cantilever on free air

Energy Lost Tapping cantilever on sample surface

Energy Lost Tapping cantilever on sample surface

ACantilever = Acantilever sin( t+ 1) 1 ACantilever(scanning) = Acantilever(Scanning) sin( 2 t+ 2)

ACantilever = Acantilever sin( t+ 1) 1 ACantilever(scanning) = Acantilever(Scanning) sin( 2 t+ 2) Tapping. Mode Cantilever oscillation amplitude in free air and during scanning.

Force gradient changes the effective resonance frequency of the cantilever In this method, the

Force gradient changes the effective resonance frequency of the cantilever In this method, the cantilever is vibrated by a small piezoelectric element near its resonant frequency. The cantilever’s resonant frequency changes in response to any additional force gradient. Attractive forces make the cantilever effectively “softer, ” reducing the cantilever resonant frequency. Conversely, repulsive forces make the cantilever effectively “stiffer, ” increasing the resonant frequency.

Force gradient changes the vibrational amplitude at the driving frequency of the cantilever A

Force gradient changes the vibrational amplitude at the driving frequency of the cantilever A (xy) is a function of Position on the sample.

Driving frequency A Phase line Resonance frequency TM AFM: ver der Waal Force gradient

Driving frequency A Phase line Resonance frequency TM AFM: ver der Waal Force gradient

Phase Imaging: Beyond Topography By mapping the phase of the cantilever oscillation during the

Phase Imaging: Beyond Topography By mapping the phase of the cantilever oscillation during the Tapping. Mode scan, phase imaging goes beyond simple topographical mapping to detect variations in composition, adhesion, friction, viscoelasticity, and perhaps other properties. Applications include identification of contaminants, mapping of different components in composite materials, and differentiating regions of high and low surface adhesion or hardness. In many cases phase imaging complements lateralforce microscopy (LFM) and force modulation techniques. often providing additional information more rapidly and with higher resolution. Phase imaging is as fast and easy to use as Tapping. Mode AFM — with all its benefits for imaging soft, adhesive, easily damaged or loosely bound samples

Force gradient changes the phase of the cantilever Phase imaging ( Elastic properties of

Force gradient changes the phase of the cantilever Phase imaging ( Elastic properties of the surface) ACantilever(scanning) = Acantilever(Scanning) sin( 2 t+ 2) x 2 y 2, x 2 y 2) x 1 y 1, x 1 y 1)

Driving frequency Phase line Resonance frequency TM AFM: ver der Waal Force gradient

Driving frequency Phase line Resonance frequency TM AFM: ver der Waal Force gradient

Condensation of Plasmid DNA into nanoparticles with Gold Nanoparticles modified with PPI G-3 Free

Condensation of Plasmid DNA into nanoparticles with Gold Nanoparticles modified with PPI G-3 Free Plasmid DNA Phase shift: 6. 1 +/- 1. 1 nm Phase shift: 0. 8 -1. 7 nm Height image Phase image

Summary In Tapping. Mode AFM, the cantilever is excited into resonance oscillation with a

Summary In Tapping. Mode AFM, the cantilever is excited into resonance oscillation with a piezoelectric driver. The oscillation amplitude is used as a feedback signal to measure topographic variations of the sample. In phase imaging, the phase lag of the cantilever oscillation, relative to the signal sent to the cantilever’s piezo driver, is simultaneously monitored. The phase lag is very sensitive to variations in material properties such as adhesion and viscoelasticity.

Need Imaging in Fluid to Enhance Resolution or In-situ Monitoring Biological or Chemical Processes

Need Imaging in Fluid to Enhance Resolution or In-situ Monitoring Biological or Chemical Processes

Electric Force Microscopy (EFM) EFM is used to map the vertical (z) and near-vertical

Electric Force Microscopy (EFM) EFM is used to map the vertical (z) and near-vertical gradient of the electric field between the tip and the sample versus the in-plane coordinates x and y. This is done using Lift. Mode. TM. The field due to trapped charges—on or beneath the sample surface—is often sufficiently large to generate contrast in an EFM image. Otherwise, a field can be induced by applying a voltage between the tip and the sample. The voltage may be applied directly from the microscope’s electronics under AFM software control, or from an external power supply with appropriate current-limiting elements in place. EFM is performed in one of three modes: amplitude detection, phase detection, or frequency modulation (FM).

Electrostatic Interactions is long range interaction

Electrostatic Interactions is long range interaction

EFM/MFM Piezoelectric materials Conductive Domains Magnetic tip, instead of only a conductive tip

EFM/MFM Piezoelectric materials Conductive Domains Magnetic tip, instead of only a conductive tip

In this method, the cantilever is vibrated by a small piezoelectric element near its

In this method, the cantilever is vibrated by a small piezoelectric element near its resonant frequency. The cantilever’s resonant frequency changes in response to any additional force gradient. Attractive forces make the cantilever effectively “softer, ” reducing the cantilever resonant frequency. Conversely, repulsive forces make the cantilever effectively “stiffer, ” increasing the resonant frequency.

Driving frequency Phase line Resonance frequency Normal TM AFM: ver der Waal Force gradient

Driving frequency Phase line Resonance frequency Normal TM AFM: ver der Waal Force gradient EFM: lift mode: Electrostatic Force gradient

Mechanical Driven Mode V piezo = Vac sin( t+ 1) Constant drive frequency Free

Mechanical Driven Mode V piezo = Vac sin( t+ 1) Constant drive frequency Free cantilever: A 1 ( d) sin( dt)+ 1) Interacted cantilever: A 2( d) sin( d t) + 2) Normal TM AFM: ver der Waal Force gradient EFM: lift mode: Electrostatic Force gradient

Electric Force Microscopy (EFM) • • Mechanical vibration given to tip Electric field will

Electric Force Microscopy (EFM) • • Mechanical vibration given to tip Electric field will produce a change in phase of tip Measure phase lag relative to piezo drive signal. In most cases, it is necessary to apply a voltage across the tip or sample to achieve a high-quality image. • Even if a layer of insulating material covered the conductive domains, they still can be detected.

EFM/MFM Piezoelectric materials Magnetic domains Conductive Domains Magnetic tip, instead of only a conductive

EFM/MFM Piezoelectric materials Magnetic domains Conductive Domains Magnetic tip, instead of only a conductive tip