Smart Meter Verification and FDA 21 CFR part

  • Slides: 33
Download presentation
Smart Meter Verification and FDA 21 CFR part 11 Tom O’Banion Director of Industry

Smart Meter Verification and FDA 21 CFR part 11 Tom O’Banion Director of Industry Differentiation

Agenda l Review of Terms l Third Party Agency Work – Overall update –

Agenda l Review of Terms l Third Party Agency Work – Overall update – FDA l AMS Snap-on and Pro. Link advantages – AMS Screen shots

Terminology: Calibration, Validation, and Verification l Definition of terms – Calibration: Establishing the relationship

Terminology: Calibration, Validation, and Verification l Definition of terms – Calibration: Establishing the relationship between flow and signal produced by sensor. – Validation: Confirming flow performance by comparing a primary flow standard to sensor. In F&B, Pharma, and Life. Science, Processes are also often validated – Verification: Establishing confidence in performance by analysis of secondary variables associated with flow l l l Frequently these terms are used interchangeably Many times calibration or validation is done when only verification is needed Micro Motion measures tube stiffness to verify complete performance

Agency Application SMV Status AER (Alberta) HC fiscal transfer ü Recognized US EPA Green.

Agency Application SMV Status AER (Alberta) HC fiscal transfer ü Recognized US EPA Green. House Gas (40 CFR part 98) ü Recognized US EPA Acid Rain (40 CFR part 75) ü (EPA will use AGA docs, FY 15) ISO 9001 Various ü Recognized IEC (for SIS) Safety Instrumented Systems ü Recognized via Proof Test AGA -11 / API MPMS Ch 14. 9 Natural Gas fiscal transfer ü Recognized Nmi (NEL and Pigsar test labs) Test data used for other Agencies ü Q 2 2015 Russia Various metrology Est Sep/Oct 2015 NIM (China) CT, Various JJG 1038 -2008 rev? Staring April 2014 FDA & TTB Meter validation ü SMV complies w/ proper recordkeeping API 20. 2 Allocation / royalty meters Ballot Feb 2015, est Sep 2015 release API 5. 6 Fiscal Transfer Liquid HC TBD 2016 NIST Liquid chemical fiscal transfer Not yet active

üISO 9001 (Often used in F&B) 7. 6 Control of monitoring and measuring equipment

üISO 9001 (Often used in F&B) 7. 6 Control of monitoring and measuring equipment The organization shall determine the monitoring and measurement to be undertaken and the monitoring and measuring equipment needed to provide evidence of conformity of product to determined requirements. The organization shall establish processes to ensure that monitoring and measurement can be carried out and are carried out in a manner that is consistent with the monitoring and measurement requirements. Where necessary to ensure valid results, measuring equipment shall a) be calibrated or verified, or both, at specified intervals, or prior to use, against measurement standards traceable to international or national measurement standards; where no such standards exist, the basis used for calibration or verification shall be recorded (see 4. 2. 4); b) be adjusted or re-adjusted as necessary; c) have identification in order to determine its calibration status; d) be safeguarded from adjustments that would invalidate the measurement result; e) be protected from damage and deterioration during handling, maintenance and storage. In addition, the organization shall assess and record the validity of the previous measuring results when the equipment is found not to conform to requirements. The organization shall take appropriate action on the equipment and any product affected. Records of the results of calibration and verification shall be maintained (see 4. 2. 4).

üFDA • Official MMI letter stating SMV is “manufacturer recommended best practice” • Part

üFDA • Official MMI letter stating SMV is “manufacturer recommended best practice” • Part of Rosemount Hygienic initiative • FAQ developed

üFDA FAQ for Sales Document # GI 002000 Rev A

üFDA FAQ for Sales Document # GI 002000 Rev A

FAQ Excerpts l l Add SMV to SOP Run SMV once per quarter and

FAQ Excerpts l l Add SMV to SOP Run SMV once per quarter and file report Share in conjunction with SMV and Zero Verification Tool videos Ask probing questions about current work practices, cost of calibration, etc. l l Meters with 800 ECP can be easily upgraded to SMV in the field When quoting a project, use the entire portfolio based on a) hygienic approvals, b) need for hygienic flanges, and c) SMV availability (e. g. Tseries is not yet SMV enabled)

Remember, Only Micro Motion Offers a True Verification Method for Coriolis: Easy, Fast, While

Remember, Only Micro Motion Offers a True Verification Method for Coriolis: Easy, Fast, While the meter continues to operate l l l Full flow meter check: tests sensor components, tube stiffness, electronics, and wiring – Compares performance and repeatability to the day meter was calibrated – Loop-back checks of electronics and signal processing to ensure total system is functioning correctly Immediate results: no need to trend and interpret results Convenient: Test can be done with several interfaces and can be scheduled Safe and Easy: Does not take meter out of the line or stop the process Fast: Test only take 90 seconds Internal/External Requirements • SMV provides traceable reporting Difficult Processes • SMV tracks meter changes to gain confidence in Flow Calibration Factor Installation Effects • Coriolis meters do not require flow conditioning Meter Degradation • Coriolis has no moving parts

SMV You. Tube video Smart Meter Verification video: https: //www. youtube. com/watch? v=Kmp 7

SMV You. Tube video Smart Meter Verification video: https: //www. youtube. com/watch? v=Kmp 7 e. Za. F 3 KM Zero Verification video: https: //www. youtube. com/watch? v=d. Y_ENu. E 1 H 1 k

SMV & AMS

SMV & AMS

Additional Benefits w/ AMS & Prolink l Enables complete test traceability – Captures meter

Additional Benefits w/ AMS & Prolink l Enables complete test traceability – Captures meter identification, configuration, and fingerprint – Test conditions recorded for user, date, time and process conditions l Scheduling – Test automatically runs on user defined interval – Print or Save a Report of Results SMV Test via Prolink SMV Report

Meter Verification SNAP-ON

Meter Verification SNAP-ON

Launch Meter Verification From Device Menu

Launch Meter Verification From Device Menu

Reading Sensor Information

Reading Sensor Information

Enter Test Information

Enter Test Information

Start Meter Verification Test

Start Meter Verification Test

Alternate Older transmitters may have limited options for Output Behavior

Alternate Older transmitters may have limited options for Output Behavior

Running Test Progress

Running Test Progress

Saving Meter Verification Data

Saving Meter Verification Data

Test Results are Displayed

Test Results are Displayed

Click Next for Detailed Results

Click Next for Detailed Results

Display or Print Full Report

Display or Print Full Report

Results Documented in Audit Trail

Results Documented in Audit Trail

Meter Verification Scheduler

Meter Verification Scheduler

Launch Scheduler From Tools Menu

Launch Scheduler From Tools Menu

Enter Optional Information Report Settings Tester Information Email Settings

Enter Optional Information Report Settings Tester Information Email Settings

Add Task

Add Task

Set Task Properties and Frequency

Set Task Properties and Frequency

Task Is Now Scheduled

Task Is Now Scheduled

Example AMS. OPCServer Data

Example AMS. OPCServer Data

What To Look For l enable_meter_verification – Provides the current meter verification test state

What To Look For l enable_meter_verification – Provides the current meter verification test state – 0 = idle, disabled, abort – 1 = normal verification (seen when hold last measurement or fault is selected) – 6 = background meter verification (seen when continue measurement is selected) l device_specific_status_4 – Provides meter verification test status – 0 = None, pass – 1 = A 032, meter verification in progress, output fixed

Committed to solving your toughest challenges

Committed to solving your toughest challenges