SAFE USE OF DIATHERMY RACHEAL AFOLABI RN RM
SAFE USE OF DIATHERMY • RACHEAL AFOLABI, RN, RM, H. Dip. PERI -OPERATIVE NURSING, H. Dip. HEALTH SYSTEM MANAGEMENT
Diathermy is one of the most commonly used tool in the operating theatre
DEFINITION • Diathermy is the generation of heat in body tissues by means of radiofrequency energy (passage of an electric current through the tissues). The energy is used to cut tissues, seal blood vessels and destroy unwanted cells by the deliberate application of intense heat.
Two basic patient circuits are used • Monopolar • Bipolar
MONOPOLAR • During monopolar diathermy the patient forms part of the electrical circuit, only one side of the generator output is connected to the active electrode. The other side is connected to a large patient return plate. The plate provides a safe path for the current to pass back to the diathermy.
Completed Circuit
BIPOLAR • During bipolar diathermy the current flows between the tips of the forceps, one tip acting as an active electrode and the other as a return. The patient’s tissue acts as the conducting medium, through which the current will flow. One advantage with this type is that production of the cutting current is virtually impossible. The field of coagulation is limited to the contact area;
BIPOLAR (Cont. ) • the surrounding tissues are not damaged. There is no patient plate attached.
EFFECTS OF SURGICAL DIATHERMY • CUTTING • COAGULATION/DESICCATION • FULGURATION
PATIENT PLATE/RETURN ELECTRODE • A minimum conductive area of 10 sq inches is recommended. Plate must be kept in secure contact with the patient’s body surface.
SETTINGS ON ELECTROSURGICAL UNIT • • Blends setting Autocut Autocoag Users must ensure that they are familiar with all settings and their functions
TYPES OF PATIENT PLATES • REUSABLE • DISPOSABLE
DISPOSABLE PLATES • CONDUCTIVE ADHESIVE PLATES - Generally thin, lightweight & flexible - Conductive adhesive which works like gel - Continuous adhesive border
ROUTINE DAILY THEATRE CHECK • This must include checking the electrosurgical unit to ensure that: - All cables & electrodes insulation are intact - Adequate supply of patient plates - Functioning alarm systems - Connections are all fitting securely - Foot pedals are all present & working
THINK SAFETY!!!
POINTS TO REMEMBER IN PROPER PATIENT PLATE USE • Ensure regular, planned preventative maintenance checks • Any faulty equipment must be removed immediately • Plates are single use • Avoid placement near scars, implant sites or ECG electrodes • A muscular well vascularised area is most suitable
POINTS (Cont. ) • Site must be clean, dry & shaved • Protect skin integrity by ensuring pt is not resting on dispersive plate clamp • The diathermy setting is determined by the surgeon • Be familiar with normal power settings • Do not allow fluid to pool at dispersive site • Check pt contact & connection before commencing
POINTS (Cont. ) • Ensure pt is not in contact with any metal • Place active electrode in a quiver when not in use • Only aqueous fluids should be used for irrigation • On completion of procedure, remove the plate carefully & inspect the skin • Document use of diathermy in pt’s record
SUGGESTED SITES FOR PLATE PLACEMENT • • • CALF UPPER ARM ABDOMEN MID BACK BUTTOCKS ANTERIOR & POSTERIOR THIGH
THINK SAFETY!!!
Pad Site Burn AORN Journal, March 1991
HAZARDS OF ELECTROSURGERY • • ACCIDENTAL BURNS SURGICAL SMOKE MINIMAL ACCESS SURGERY ENDOSCOPIC SURGERY
CAUSES OF THERMOELECTRIC BURNS • Poor patient plate contact • Hair at the site of return electrode plate • Break in the wire connecting the plate to the electro-surgical unit • Detached return cable • Ignition of flammable material on pt’s skin • High voltage electrocution of the pt • Abnormally high power setting
References • 3 M Health Care (2000) Safety in Surgical Diathermy 3 M Health Care Ltd. , England • Wicker, P. (2000) Electrosurgery in perioperative practice. British Journal of Perioperative Nursing. Vol 10 (4).
Cont…. . • Electrosurgical Safety Guide, Con. Med Corporation, USA.
THANK YOU FOR LISTENING