Rotating Coil System Requirements and Specs J Di

  • Slides: 2
Download presentation
Rotating Coil System Requirements and Specs. J. Di. Marco 08 Jun 2018 mag. Meas.

Rotating Coil System Requirements and Specs. J. Di. Marco 08 Jun 2018 mag. Meas. Sys. Review_RCreq_jdm_08 Jun 2018 1

As stated in US-Hi. Lumi-doc-818 Rotating Coil Measurements for MQXF are required to measure

As stated in US-Hi. Lumi-doc-818 Rotating Coil Measurements for MQXF are required to measure the following : 1. 2. 3. 4. Integrated quadrupole field with accuracy better than 0. 1% and resolution better than 0. 01% Integral harmonics End field harmonics determined with z increments < 0. 25 m axially Harmonic coefficients accuracy of 5 ppm (10 ppm) of main field in the body region for orders 3 to 10 (11 to 14) at the reference radius of 50 mm. 5. Local twist variation of the quadrupole roll angle as a function of axial position at the 0. 5 mrad level 6. Local magnetic axis (‘waviness’) with respect to external fiducials at the 0. 2 mm level With the present magnetic measurement system at VMTF, a short probe is scanned with accurate Zincrements along the magnet axis. This provides both local and integral data (via summation). Results comparable to Requirements 1 -4 are routinely achieved during these measurements. Ø With a comparable z-positioning accuracy and comparable probe system, Req. 1 -4 should be straightforward to achieve for the full-length Q 1/Q 3 cryo-assembly Ø Req. 5 -6 will depend on determining 3 -D probe location in magnet during changes in Z position mag. Meas. Sys. Review_RCreq_jdm_08 Jun 2018 2