Methanol and Ethanol Analysis The Ultimate Gourmet Martini

  • Slides: 14
Download presentation
Methanol and Ethanol Analysis/ The Ultimate Gourmet Martini J. Picker HP Labs Escondido, CA

Methanol and Ethanol Analysis/ The Ultimate Gourmet Martini J. Picker HP Labs Escondido, CA

Objective n n Analyze methanol and ethanol at 100 -1000 ug/L or lower Use

Objective n n Analyze methanol and ethanol at 100 -1000 ug/L or lower Use standard or modified EPA Methods Provide adequate Turnaround and pricing to meet client needs Make a profit HP Labs Escondido, CA

Warning n n Methanol is “wood alcohol” minimum lethal dose 1 g/kg but toxic

Warning n n Methanol is “wood alcohol” minimum lethal dose 1 g/kg but toxic effects as low as 10 g. toxicity is due to metabolic production of formic acid Ethanol is less toxic HP Labs Escondido, CA

Problems n n Very Poor purging efficiency (i. e. , NO 5030/5035) Allowed Prep.

Problems n n Very Poor purging efficiency (i. e. , NO 5030/5035) Allowed Prep. Techniques 5031 Azeotropic distillation u 5032 Vacuum distillation u 5021 Equilibrium Headspace =nd u 5041 VOST train = nd u Direct Inject u HP Labs Escondido, CA

Problems II n n Low sensitivity on FID and MS Interferences by HCs and

Problems II n n Low sensitivity on FID and MS Interferences by HCs and other compounds on FID HP Labs Escondido, CA

HP Labs’ Solution n n Direct Aqueous injection/EPA Method 8260 B GC Column and

HP Labs’ Solution n n Direct Aqueous injection/EPA Method 8260 B GC Column and conditions to separate Me. OH, Et. OH from water Stable Labeled isotope internal standards (CD 3 OH, CD 3 CH 2 OH) Modified with Selected Ion Monitoring Mass Spectrometry (SIM) HP Labs Escondido, CA

HP Labs Escondido, CA

HP Labs Escondido, CA

HP Labs Escondido, CA

HP Labs Escondido, CA

HP Labs Escondido, CA

HP Labs Escondido, CA

HP Labs Escondido, CA

HP Labs Escondido, CA

Sample Preparation: water n n n Sample Size 1 m. L Add Labeled IS

Sample Preparation: water n n n Sample Size 1 m. L Add Labeled IS Solution Inject HP Labs Escondido, CA

Sample Preparation: Soil n n Ultrasonic extraction (EPA 3550) 10 g/10 m. L Remove

Sample Preparation: Soil n n Ultrasonic extraction (EPA 3550) 10 g/10 m. L Remove 1 m. L Add IS Inject HP Labs Escondido, CA

Limits n n Detection Water and Soil MDLs < 100 ug/L (ug/kg) EQLs 250

Limits n n Detection Water and Soil MDLs < 100 ug/L (ug/kg) EQLs 250 ug/L (ug/kg) HP Labs Escondido, CA

What This Means n n Methanol and Ethanol on site or off site Rapid

What This Means n n Methanol and Ethanol on site or off site Rapid turnaround All levels of QC Price as low as fixed base HP Labs Escondido, CA