Luminosity Monitoring For Prim Ex Dan Dale Prim

  • Slides: 11
Download presentation
Luminosity Monitoring For Prim. Ex Dan Dale Prim. Ex Collaboration Meeting February 23, 2007

Luminosity Monitoring For Prim. Ex Dan Dale Prim. Ex Collaboration Meeting February 23, 2007

E. P. point 14 • “Electron counting systematics listed as 0. 8%. How did

E. P. point 14 • “Electron counting systematics listed as 0. 8%. How did you get this number? ” Electron Counting Systematics 0. 8% discrepancy between CLOCK and OUT-OF-TIME methods

E. P. point 15 • “You discuss in the Appendix effects of beam intensity

E. P. point 15 • “You discuss in the Appendix effects of beam intensity instabilities on rate sampling. You don’t quantify this effect and do not include in final error” 0. 8% discrepancy between CLOCK and OUT-OF-TIME methods CLOCK trigger sampling used -> insensitive to beam instabilities

Point 11: Clock frequency, window width • Ne = ne / (W x ntrigs)

Point 11: Clock frequency, window width • Ne = ne / (W x ntrigs) Live 1 • Clock frequency = 195. 3125 +-0. 020 k. Hz • W = 7 microseconds – integral linearity < 25 ppm differential linearity 100 psec

DS point: Different PS dipole settings--run 4339 collimator 0. 1 “ off

DS point: Different PS dipole settings--run 4339 collimator 0. 1 “ off

Effect of Collimator Position on Tagging Ratios

Effect of Collimator Position on Tagging Ratios

Effect of Collimator Position on Tagging Ratios – units are inches

Effect of Collimator Position on Tagging Ratios – units are inches

TAC losses 0. 1% interact in vacuum window & Helium – overwhelmingly pair production

TAC losses 0. 1% interact in vacuum window & Helium – overwhelmingly pair production

Point 13: “Section 25 all numbers in two tables identical except last one. Something

Point 13: “Section 25 all numbers in two tables identical except last one. Something is not right” Luminosity Error Budget ____________________ effect error(%) Uncorrected leakage current 0. 5 (1. 4%/3) TAC statistics 0. 2 (0. 5 per T channel) Electron counting systematics 0. 8 (clock versus out of time) Electron counting statistics negligible TAC reproducibility 0. 4 HYCAL collimation of TAC 0. 25 (worst case, no photon collimator) Beam position/collimator effect 0. 2 Absorption in TAC runs not yet evaluated ________________________________ 1. 1%

What we can ultimately do. Luminosity Error Budget ____________________ effect error(%) Corrected leakage current

What we can ultimately do. Luminosity Error Budget ____________________ effect error(%) Corrected leakage current 0. 3 (1. 4%/3/2) TAC statistics 0. 2 (0. 5 per T channel) electron counting systematics 0. 8 (clock versus out of time) Electron counting statistics negligible TAC reproducibility 0. 4 HYCAL collimation of TAC 0. 25 (worst case, no photon collimator) Beam position/collimator effect 0. 2 Absorption in TAC runs not yet evaluated ________________________________ 1. 0%

 • Points 1, 3: “Please define meaning of Tcounter and T channel” •

• Points 1, 3: “Please define meaning of Tcounter and T channel” • Point 2, 16: “In section 4 (2002 run analysis) just T counters used. E-T coincides since implemented. Needs to be written up. • Point 5: absolute efficiency of TAC – dedicated run? ?