LP 20 e ORIENTATION GUIDE Youre Ready 2010

  • Slides: 30
Download presentation
LP 20 e ORIENTATION GUIDE You’re Ready. © 2010 Physio-Control, Inc. All rights reserved.

LP 20 e ORIENTATION GUIDE You’re Ready. © 2010 Physio-Control, Inc. All rights reserved.

Important Safety Information This orientation program is not a substitute for hands-on device training.

Important Safety Information This orientation program is not a substitute for hands-on device training. The LIFEPAK 20 e defibrillator is intended for use by personnel who have been trained in the use of the LIFEPAK 20 e defibrillator © 2010 Physio-Control, Inc. All rights reserved.

Objectives Review the operation and features of the LIFEPAK 20 e defibrillator including: §

Objectives Review the operation and features of the LIFEPAK 20 e defibrillator including: § § § Electrode pad placement Analyse & manual mode Cardiac pacing Synchronised cardioversion Troubleshooting Maintenance © 2010 Physio-Control, Inc. All rights reserved.

LIFEPAK 20 e Defibrillator Uses: § Manual defibrillation - For use only on patients

LIFEPAK 20 e Defibrillator Uses: § Manual defibrillation - For use only on patients in cardiopulmonary arrest § Semi-automated external defibrillation - For use only on patients in cardiopulmonary arrest § ECG monitoring - For the purpose of ECG rhythm recognition and heart rate monitoring § External cardiac pacing § Synchronised cardioversion © 2010 Physio-Control, Inc. All rights reserved.

Get to know the device On/Off button Analyzing button Shock button © 2010 Physio-Control,

Get to know the device On/Off button Analyzing button Shock button © 2010 Physio-Control, Inc. All rights reserved.

Basic steps for manual or automated use § Establish patient is in cardiac arrest

Basic steps for manual or automated use § Establish patient is in cardiac arrest § Dial 2222/9999 § Turn on the LP 20 e § Apply electrodes § Confirm rhythm or § Follow screen messages and voice prompts © 2010 Physio-Control, Inc. All rights reserved.

Electrode placement § § Remove clothing from patient’s chest Remove excessive hair Clean and

Electrode placement § § Remove clothing from patient’s chest Remove excessive hair Clean and dry the skin Do not apply alcohol, tincture of benzoin, or antiperspirant to the skin © 2010 Physio-Control, Inc. All rights reserved.

Electrode placement Press electrodes firmly onto patient’s skin © 2010 Physio-Control, Inc. All rights

Electrode placement Press electrodes firmly onto patient’s skin © 2010 Physio-Control, Inc. All rights reserved.

Electrode placement Correct electrode position optimizes the amount of current flowing through the ventricles

Electrode placement Correct electrode position optimizes the amount of current flowing through the ventricles Correct electrode position Incorrect electrode position © 2010 Physio-Control, Inc. All rights reserved.

ECG analysis in Analyse mode § When an analysis occurs, the defibrillator will prompt

ECG analysis in Analyse mode § When an analysis occurs, the defibrillator will prompt to stand clear of the patient § Do not touch the patient during analysis § Make sure everyone is clear of the patient © 2010 Physio-Control, Inc. All rights reserved.

Shock Advised! § If the defibrillator has detected a shockable ECG rhythm § The

Shock Advised! § If the defibrillator has detected a shockable ECG rhythm § The LP 20 e will charge automatically in analyse mode. § In manual mode you would have to select the charge button once the door is opened © 2010 Physio-Control, Inc. All rights reserved.

Shock Advised! § Continue chest compressions whilst the device is charging! § Clear everyone

Shock Advised! § Continue chest compressions whilst the device is charging! § Clear everyone from the patient or any equipment touching the patient § Press the SHOCK button © 2010 Physio-Control, Inc. All rights reserved.

CPR time q In manual mode you are required to start CPR immediately q

CPR time q In manual mode you are required to start CPR immediately q Start CPR prompt occurs immediately after shock is delivered in Analyse mode q A text message and countdown timer appear for the remaining CPR time q At the end of CPR time, the defibrillator will request you to push ANALYSE © 2010 Physio-Control, Inc. All rights reserved.

Shock Advised Continue to follow prompts until patient has return of spontaneous circulation or

Shock Advised Continue to follow prompts until patient has return of spontaneous circulation or until the cardiac arrest/medical emergency team arrive. © 2010 Physio-Control, Inc. All rights reserved.

No Shock Advised (In Analyse mode) § § § The LP 20 e has

No Shock Advised (In Analyse mode) § § § The LP 20 e has determined the heart rhythm is not shockable The LP 20 will prompt to ‘check for pulse’: if no pulse start CPR Perform CPR until the patient has return of spontaneous circulation or the LP 20 e requests to reanalyse the heart rhythm © 2010 Physio-Control, Inc. All rights reserved.

Return of spontaneous circulation § Position the patient according to training if consciousness and/or

Return of spontaneous circulation § Position the patient according to training if consciousness and/or signs of circulation and breathing return § Leave electrodes attached § Keep patient as still as possible § Electrodes may be connected to another LIFEPAK defibrillator © 2010 Physio-Control, Inc. All rights reserved.

Battery life of the LP 20 e § Charge level is indicated on the

Battery life of the LP 20 e § Charge level is indicated on the main screen with a green battery icon with bars § Fully charged battery: all green lights illuminated § On a full battery you should have sufficient life to transport a patient for up to 4 hours. If defibrillating this will significantly reduce and should be plugged in mains whenever possible. © 2010 Physio-Control, Inc. All rights reserved.

Battery § Low battery: single steady green light. § Very low battery: battery icon

Battery § Low battery: single steady green light. § Very low battery: battery icon will flash § Battery icon location © 2010 Physio-Control, Inc. All rights reserved.

Pacing q Ensure the ECG cable is connected and placed on lead II q

Pacing q Ensure the ECG cable is connected and placed on lead II q Apply defibrillator pads to patient’s bare chest © 2010 Physio-Control, Inc. All rights reserved.

Pacing q Activate the pacer q The device is preset to 60 BPM on

Pacing q Activate the pacer q The device is preset to 60 BPM on demand q Increase the current as trained on ALS course © 2010 Physio-Control, Inc. All rights reserved.

Synchronised cardioversion q Defibrillator pads must be applied to the patient q Press the

Synchronised cardioversion q Defibrillator pads must be applied to the patient q Press the SYNC button q Select the correct energy for defibrillation q The ‘R’ wave will be detected by at above the ECG © 2010 Physio-Control, Inc. All rights reserved.

Troubleshooting tips § Message: ‘Connect electrodes’ § ‘Check connector and electrodes’ § Possible causes:

Troubleshooting tips § Message: ‘Connect electrodes’ § ‘Check connector and electrodes’ § Possible causes: § Electrodes are not adhered to the patient’s skin § Electrodes are disconnected from the LP 20 e § Patient impedance measurement outside of range © 2010 Physio-Control, Inc. All rights reserved.

Troubleshooting tips § Message: ‘Connect electrodes’ § ‘Check connector and electrodes’ § What to

Troubleshooting tips § Message: ‘Connect electrodes’ § ‘Check connector and electrodes’ § What to do § Firmly press electrodes onto skin § Check that connector is completely inserted into the LP 20 e lead § Clean, shave and dry skin § Replace electrode pads § If prompt continues, perform CPR until ALS arrives © 2010 Physio-Control, Inc. All rights reserved.

Troubleshooting tips § Message: ‘Motion detected, stop motion’ § Possible causes: § Patient movement

Troubleshooting tips § Message: ‘Motion detected, stop motion’ § Possible causes: § Patient movement due to: breathing, CPR being performed during analysis, some internal pacemakers, vehicle motion, or electrical/radio frequency interference § What to do: § § Check patient for normal breathing Stop CPR during analysis Stop vehicle during analysis Move communication or other suspected devices away from the LP 20 e when possible © 2010 Physio-Control, Inc. All rights reserved.

Maintenance q Daily test § Connect the end test adapter into therapy cable and

Maintenance q Daily test § Connect the end test adapter into therapy cable and select: § Options § User test § Push jog shuttle to confirm § And confirm you wish to continue q The device will perform a test and produce a test strip to retain for 6 months q This must be performed ‘SHIFTLY © 2010 Physio-Control, Inc. All rights reserved.

Loading 50 mm paper The printer is equipped with an out-of-paper sensor to protect

Loading 50 mm paper The printer is equipped with an out-of-paper sensor to protect the print-head. The sensor automatically turns off the printer if paper runs out or if the printer door is open. To load the paper: 1 Press the black button to open the printer door. 2 Remove the empty paper roll. 3 Insert the new paper roll, grid facing upward. 4 Pull out a short length of paper. 5 Push the printer door in to close. Paper reload © 2010 Physio-Control, Inc. All rights reserved.

Maintenance § If your device requires service then, this must be done through the

Maintenance § If your device requires service then, this must be done through the estates helpdesk… © 2010 Physio-Control, Inc. All rights reserved.

After-use care After the LP 20 e has been used: § § § Clean

After-use care After the LP 20 e has been used: § § § Clean foreign substances from the LP 20 e Replace electrodes Replenish supplies as needed Ensure device is plugged into mains Spare defibrillator pads are available on Ward 3 ‘Centralised Resuscitation Store’ via the § The Duty Resuscitation Officer § The Acute Response Team © 2010 Physio-Control, Inc. All rights reserved.

Routine maintenance § The LP 20 e is easy to maintain § On a

Routine maintenance § The LP 20 e is easy to maintain § On a periodic basis check the following: § There are no signs of wear § The device can hold a battery charge if taken off the mains § The Use-By date on electrode pads is not passed (must be checked as part of your shiftly check) § Check for damage to or foreign substances on the LP 20 e © 2010 Physio-Control, Inc. All rights reserved.

Thank you © 2010 Physio-Control, Inc. All rights reserved.

Thank you © 2010 Physio-Control, Inc. All rights reserved.