Introducing the Varian 356 LC Refractive Index Detector




























- Slides: 28

Introducing the Varian 356 -LC Refractive Index Detector

Introduction The Varian 356 -LC detector is being introduced to replace the Pro. Star 355 detector. The 356 -LC detector was designed and built at Polymer Labs, UK. The Varian 356 -LC is a high performance refractive index detector designed to meet virtually all customer needs for refractive index detection. This universal detector is ideal for the analysis of non-absorbing compounds such as polymers, carbohydrates, alcohols, and fatty acids. 2

Overview The Varian 356 -LC has been designed to integrate into the Pro. Star range of detectors. 296 mm Status LED Showing Power, Purge, zero and error 211 mm On/Off switch 477 mm Access panel to connections 3 Hinges for “Barbie” shower Leak drain

Overview Outlet Waste for internal leaks 4 Inlet

Principle of Operation Refractive index detectors are universal detectors designed for analysis of compounds that do not strongly absorb UV light and, therefore, cannot be assayed with a UV detector. The usual compounds analyzed are sugars, polymers, aliphatic amines, organic acids, and alcohols. The Varian 356 -LC is functionally identical to other deflection type refractive index detector, it comprises: • LED light source, • 6 µL flowcell with reference and sample compartments • split photodiode detector The complete assembly is housed in a thermally controlled block to maintain baseline stability. 5

Principle of Operation Generic diagram of the Varian 356 -LC Refractive Index Detector 6

Principle of Operation I 1 I 2 When both cell compartments contain the same solvent, the light beam is positioned equally on the two detector diodes I 1 and I 2 When solvent containing a compound passes through the sample side of the cell the difference in refractive index causes the light beam to be proportionally deflected (according to Snell’s refraction law). The change in light intensities between I 1 & I 2, is proportional to concentration and refractive index of the sample solution. Refractive index is a bulk property of the mobile phase and is very sensitive to changes in ambient temperature RI detectors are also sensitive to flow-rate changes and cannot be used for gradient elution. 7

Internal Flow Path Normal Operating Conditions: • The incoming solvent passes into & out of the sample side of the flowcell, then to waste Purge Conditions: • The incoming solvent passes into sample & reference side of the flowcell, before going to waste The Varian 356 -LC has only 15 µL of volume between inlet and cell to keep dispersion to a minimum The maximum recommended flow rate permitted through the Varian 356 -LC Detector is 10 m. L/min 8

Operational Parameters Purge When Purge is selected, the purge indicator light is highlighted, and solvent flow is diverted through the reference side of the RI cell, instead of the sample side; before going to waste. The purge time can be adjusted to compensate for different solvents & eluent flow rates. Auto-Zero The detector output is re-set to zero when the auto-zero function is selected. An error will occur if the auto-zero process takes longer than 5 minsn Response Time The Varian 356 -LC has a default response time of 1. 0 second, with a settable range of 0. 1 – 5. 0 seconds. 9

Operational Parameters Sensitivity The sensitivity of the Varian 356 -LC can be adjusted to provide greater flexibility. The three possible sensitivity settings are described below: • Low - Suitable for highly concentrated samples (e. g. Prep LC) • Standard - Default setting, ideal for the majority of LC applications • High - Suitable for samples of low concentration in order to maximize signal‑to‑noise. Temperature The temperature of the Varian 356 -LC can be set between 30 -50 °C (1 °C increments) or turned off. The default value is 35 °C. It is recommended that the temperature should be set at least 5 °C above ambient temperature in order to ensure a stable thermal environment. 10

Operational Parameters Polarity The output of the Varian 356 -LC refractive index detector can be either positive or negative; therefore it is essential that the polarity of the detector signal be configured. The polarity parameter is used to invert the sign of the refractive index data. A positive polarity leaves the data unaffected, whilst a negative polarity essentially inverts the chromatogram. Baseline Offset The baseline offset of the Varian 356 -LC can be set between the ranges – 100 to +100 µRIU. 11

Front Panel Status The Varian 356 -LC, like other Varian modules, has no front-panel display or hard-key control. Full functionality is provided by Galaxie CDS, which gives the user greater flexibility and control than the Pro. Star 355. Three indicator lights are located at the top left of the front panel of the module to display the status of the instrument 12 Status Color Ready Not Ready (i. e. changing temp) Error Green Orange Red Auto-zero on Green Purge on Green

Galaxie Control The Varian 356 -LC can be connected directly to the PC via an RS 232 connection, for full Galaxie Control Detector output and status quickly accessed in overview window Status window displays all current detector conditions A/Z & purge function easily accessed 13

Galaxie Control All detector parameters can be changed via the configuration window Send button downloads settings to detector 14

Key Benefits Sensitivity The Varian 356 -LC provides three sensitivity settings, HIGH, STD & LOW, that can be selected according to the type of analysis. For example, the LOW setting is ideal for preparative LC in order to prevent the signal from saturating at the high concentrations typically encountered. Baseline Stability Flow cell thermostating in the Varian 356 -LC eliminates temperature effects on the RI baseline. Stable baselines mean lower detection limits for both analytical and preparative applications. Temperatures are controlled in one degree increments from 30 °C to 50 °C. Superior Peak Shape The custom designed 6µL ‘flow-efficient’ cell for minimal peak dispersion delivers outstanding peak shape, and superior resolution. 15

Key Benefits: Superior Peak Shape At 1 m. L/min the Varian 356 -LC peak width* is 50% narrower than the Pro. Star 355 Varian 356 -LC Pro. Star 355 16 *measure at ½ height

Key Benefits: Superior Resolution The Varian 356 -LC provides better resolution between peaks than the Pro. Star 355. Resolution 1. 17 This is beneficial when analysing compounds of similar properties under isocratic conditions. Resolution 1. 18 17

Key Benefits: Automating the 356 -LC Fully Automated via Galaxie Software from Varian provides control and automation capabilities. The detector can be automatically purged prior to analysis for overnight operation. The Varian 356 -LC can be externally controlled, permitting a device (for instance, a pump) to automatically activate functions such as electronic optical zero and reference cell flush. Digital Data Acquisition Unlike the Pro. Star 355 the Varian 356 -LC offers the user a digital data signal, which can be taken directly into a PC, using an RS 232 cable, without the need for an MIB 800 module. 18

Automation via Galaxie Software Detector settings can be saved in a Galaxie method along with pump program for individual applications Automatic Purge and A/Z are possible within a Galaxie method Varian 356 -LC can be automated using a sequence of Galaxie methods 19

Applications: Carbohydrates in Fruit Juice A cost-effective and simple separation to identify food adulteration in fruit juice can be achieved using the Varian 356 -LC and PL-Hiplex columns. 20

Applications: Maltodextrin in Sports Drink The low dispersion characteristics of the Varian 356 -LC provides excellent resolution for maltodextrin analysis in sports beverages. 21

Applications: Epoxy Resins The Varian 356 -LC is ideal for GPC analysis, when combined with Galaxie and Cirrus GPC software. 22

Product Positioning The Varian 356 -LC and PL-ELS 2100 ELSD are both universal detectors and should have distinct positions in the market place: ELSDs are typically more expensive than RID, and there is no justification for the higher cost and added complexity of method development if the RI detector can already adequately perform the analysis. The Varian 356 -LC should be positioned as a high performance general use refractive index detector with low noise, excellent thermal control & advanced capabilities such as automatic purge & auto zero using Galaxie software. 23

Varian 356 -LC Features & Benefits Feature Benefit Low dead-volume flow-cell Improved resolution and sensitivity Active temperature compensation Lower base-line drift and improved reproducibility Galaxie Control Improved productivity through automated, unattended sequence runs Easy to use and consistent user interface Improved traceability using Galaxie operation logs Digital data output Simple to set up Better value – do not need an expensive analog to digital interface board Long-Life LED 24 Reduced maintenance and quick start-up as the LED is always on and has a indefinite life time

Varian 356 -LC Specifications RI range (RIU) 1. 0~1. 75 Range (µRIU) 150 -600 x 10 -6 RIU/FS 200 mmm 200 m 600 Linearity (µRIU) LOW (2), STD (4), HIGH (8) Short-term noise (n. RIU) <5. 0 Drift (n. RIU/hr) <250 Response time (s) Temperature control (°C) 415 mm Sensitivity (m. V/µRIU) 0. 1– 5. 0 OFF, 30– 50 (1 °C increments) Cell volume (µL) 6. 0 Light source LED 880 nm Analog output 1 V FSD Digital output 45 405 m 0 m Flow rate range (m. L/min) m 24 bit (10 Hz) via serial port Dimensions (w x d x h) (unpacked) 296 x 475 x 212 mm PC control (software) Galaxie Software m 25 0. 1– 10

Ordering Information Varian 356 -LC Detector PL 0880 -0000 Packing List Manual for Varian 356 -LC detector 1. 5 M PTFE Outlet Tubing assembly Aux I/O Cable (5 m) RS 232 Communication cable Green Half Moon Door Cap PL 0880 -0010 PL 0880 -0300 PL 0860 -0055 PL 091 -24950 01 107288 00 09 102066 00 Accessories Analogue Cable Mains Inlet Fuses (pk 5) PL 0880 -0310 PL 0860 -0420 Field Upgrades LED assembly Flow Cell Seals (Top) Cell Seals (Bottom) Purge Valve 26 PL 0880 -0110 PL 0880 -0120 PL 0880 -0150 PL 0880 -0160 PL 0880 -0210

Product Information Folder (PIF) * The Varian 356 -LC PIF contains additional information * 27