Single vs Multirow Single-slice scanner Pitch 1. 5 Pitch 2 Blurring Pitch 3 Pitch 5 Pitch 7 Volume Zoom SURE VIEW Provided by Siemens Erlangen
„Magic Triangle“ Larger Volume Shorter Scan Time Isotropic Resolution Provided by Siemens Erlangen
Resolution þ Isotropic Imaging No isotropic voxels isotropic voxel
Speed þ 64 slices /rotation þ Rotation times < 0. 5 s possible Dynamic studies
CT -- Benchmarking þ Geometric Resolution: - xy: z: 0. 3 -0. 5 mm 0. 6 mm þ Radiometric Resolution: - 12 bit data = 4096 shades of gray Hounsefield Units (HU): Minimum: n Zeropoint: n Fat: n Bones: n -1024 HU = air 0 HU = water -120 – 80 HU > 200 HU
CT -- Benchmarking þ Time Resolution: - Standard: - Cardiac Mode: 0. 5 s 0. 33 s þ Maximal Scanlength: - ca. 150 cm
Resolution Thickness 5. 0 mm Increment 2. 5 mm Thickness 2. 5 mm Increment 2. 0 mm Thickness 1. 25 mm Increment 1. 0 mm
Resolution Thickness 5. 0 mm Increment 2. 5 mm Thickness 1. 25 mm Increment 1. 0 mm Thickness 2. 5 mm Increment 2. 0 mm
CT – New: Dual Source CT
CT & Radiaten þ CT ~ 10% of Xray Examinations þ CT source of ~50% medical related radiation þ Incresases stochastic risk for cancer development: - 1 m. Sv Totalbody irradation risk for lethal cancer is increased by 5% - USA: maybe about 500 children dying a year by cancer due to CT