First DQE measurment of 500 m thick Si

  • Slides: 30
Download presentation
First DQE measurment of 500 µm thick Si hybrid pixel sensor with Photon counting

First DQE measurment of 500 µm thick Si hybrid pixel sensor with Photon counting readout for X-ray medical imaging M. Caria, J. Chaput, L. Sarry, B. Surre IWORID 2004 -- B. SURRE Glasgow

Introduction • Aim: to measure system DQE following a standart protocole • Requirements: –

Introduction • Aim: to measure system DQE following a standart protocole • Requirements: – IEC 62220 -1 – RQA 5 (~ 70 k. Vp) – Flat field corrected data IWORID 2004 - B. SURRE 2

Establishment of Standard X-ray spectrum • Adjust k. Vp of generator to get RQA

Establishment of Standard X-ray spectrum • Adjust k. Vp of generator to get RQA 5 spectrum such as: Spectrum Added filtration (mm Al) HVL (mm Al) RQA 3 10 7 RQA 5 21 7. 1 RQA 7 30 9. 1 RQA 9 40 11. 5 IWORID 2004 - B. SURRE 3

Measurement geometry IWORID 2004 - B. SURRE 4

Measurement geometry IWORID 2004 - B. SURRE 4

Tube calibration @ RQA 5 IWORID 2004 - B. SURRE 5

Tube calibration @ RQA 5 IWORID 2004 - B. SURRE 5

Set up ? • Well known Photon counting chip = MEDIPIX 2 (chip set

Set up ? • Well known Photon counting chip = MEDIPIX 2 (chip set just above the noise) • Bump bonded with 500 µm thick Si detector Thanks to the MEDIPIX collaboration • MUROS 2 interface • Medisoft IWORID 2004 - B. SURRE 6

Depletion voltage 20 % overdepletion 120 V IWORID 2004 - B. SURRE 7

Depletion voltage 20 % overdepletion 120 V IWORID 2004 - B. SURRE 7

Conversion function • Pixel size 55 x 55 µm² • Fluence: 91 #/(pixel. µGy)

Conversion function • Pixel size 55 x 55 µm² • Fluence: 91 #/(pixel. µGy) • Slope: 5. 9 count/(pixel. µGy) • G=6. 5% IWORID 2004 - B. SURRE 8

MTF (ESF) Angle α • N = 1/tan(α) = 12 • Oversampling = 0.

MTF (ESF) Angle α • N = 1/tan(α) = 12 • Oversampling = 0. 055 mm/12 IWORID 2004 - B. SURRE 9

SNR ? • 20 flood images to compute the flat field coefficients IWORID 2004

SNR ? • 20 flood images to compute the flat field coefficients IWORID 2004 - B. SURRE 10

ESF, RQA 5, 10 m. As IWORID 2004 - B. SURRE 11

ESF, RQA 5, 10 m. As IWORID 2004 - B. SURRE 11

7. Step (300 pts), RQA 5, 10 m. As Step(x)=ESF(x-x_shift)-ESF(x) IWORID 2004 - B.

7. Step (300 pts), RQA 5, 10 m. As Step(x)=ESF(x-x_shift)-ESF(x) IWORID 2004 - B. SURRE 12

DFT(Step) (300 pts), RQA 5, 10 m. As MTF(f)=|DFT(real step, f)|/|DFT(perfect step, f)| IWORID

DFT(Step) (300 pts), RQA 5, 10 m. As MTF(f)=|DFT(real step, f)|/|DFT(perfect step, f)| IWORID 2004 - B. SURRE 13

DFT(Step) (300 pts), RQA 5, 10 m. As MTF(f)=|DFT(real step, f)|/|DFT(perfect step, f)| IWORID

DFT(Step) (300 pts), RQA 5, 10 m. As MTF(f)=|DFT(real step, f)|/|DFT(perfect step, f)| IWORID 2004 - B. SURRE 14

DFT(Step) (300 pts), RQA 5, 10 m. As MTF(f)=|DFT(real step, f)|/|DFT(perfect step, f)| IWORID

DFT(Step) (300 pts), RQA 5, 10 m. As MTF(f)=|DFT(real step, f)|/|DFT(perfect step, f)| IWORID 2004 - B. SURRE 15

MTF, RQA 5, 10 m. As IWORID 2004 - B. SURRE 16

MTF, RQA 5, 10 m. As IWORID 2004 - B. SURRE 16

MTF, RQA 5, 10 m. As IWORID 2004 - B. SURRE 17

MTF, RQA 5, 10 m. As IWORID 2004 - B. SURRE 17

MTF, RQA 5, 10 m. As Presampling resolution: 7. 7 lp/mm Visible detail limit

MTF, RQA 5, 10 m. As Presampling resolution: 7. 7 lp/mm Visible detail limit = 0. 3 IWORID 2004 - B. SURRE 18

NPS @ RQA 5 dose IWORID 2004 - B. SURRE 19

NPS @ RQA 5 dose IWORID 2004 - B. SURRE 19

NTF² @ RQA 5 IWORID 2004 - B. SURRE 20

NTF² @ RQA 5 IWORID 2004 - B. SURRE 20

DQE @ RQA 5 DQE(f) = Gx. MTF²/NTF² IWORID 2004 - B. SURRE 21

DQE @ RQA 5 DQE(f) = Gx. MTF²/NTF² IWORID 2004 - B. SURRE 21

NPS considerations & calculation • Can we consider our results correct? >>> Calculations and

NPS considerations & calculation • Can we consider our results correct? >>> Calculations and simulations… IWORID 2004 - B. SURRE 22

Experimental NPS dose IWORID 2004 - B. SURRE 23

Experimental NPS dose IWORID 2004 - B. SURRE 23

Simulated NPS dose IWORID 2004 - B. SURRE 24

Simulated NPS dose IWORID 2004 - B. SURRE 24

NPS comparison: sim & exp IWORID 2004 - B. SURRE 25

NPS comparison: sim & exp IWORID 2004 - B. SURRE 25

1 -D NPS comparison, 2 directions IWORID 2004 - B. SURRE 26

1 -D NPS comparison, 2 directions IWORID 2004 - B. SURRE 26

1 -D NPS comparison IWORID 2004 - B. SURRE 27

1 -D NPS comparison IWORID 2004 - B. SURRE 27

1 -D NPS comparison Interpolation increases the low frequencies IWORID 2004 - B. SURRE

1 -D NPS comparison Interpolation increases the low frequencies IWORID 2004 - B. SURRE 28

Conlusions • DQE has been calculated on 500 µm thick Si, bonded to a

Conlusions • DQE has been calculated on 500 µm thick Si, bonded to a MEDIPIX 2 chip • A standard measure protocol have been used • DQE is not so evident to find… • How to improve this DQE ? (remember medicine) • Next step: comparison with detectors of different thickness, study in parallel with detectors from the market, further calculations… IWORID 2004 - B. SURRE 29

Acknowledgements • The MEDIPIX collaboration • J. CHAPUT, M. CARIA • L. SARRY •

Acknowledgements • The MEDIPIX collaboration • J. CHAPUT, M. CARIA • L. SARRY • You, for your attention IWORID 2004 - B. SURRE 30