Figures for Chapter 4 Electroacoustic Performance Dillon 2001

  • Slides: 23
Download presentation
Figures for Chapter 4 Electroacoustic Performance Dillon (2001) Hearing Aids

Figures for Chapter 4 Electroacoustic Performance Dillon (2001) Hearing Aids

Ear simulator V 1 V 2 V 3 V 4 Microphone Dampers Figure 4.

Ear simulator V 1 V 2 V 3 V 4 Microphone Dampers Figure 4. 1 Simplified internal structure of a four-branch ear simulator. Source: Dillon (2001): Hearing Aids

Couplers and ear simulators Photo removed to minimize file space Figure 4. 2 Several

Couplers and ear simulators Photo removed to minimize file space Figure 4. 2 Several couplers and their adapters, and an ear simulators. Source: Dillon (2001): Hearing Aids

2 -cc couplers ITE / ITC / CIC Figure 4. 3 The internal dimensions

2 -cc couplers ITE / ITC / CIC Figure 4. 3 The internal dimensions and coupling methods for several 2 -cc couplers. Putty HA 1 Microphone 2 mm dia Insert earphone 25 Earmold simulator 18 mm 18 3 mm dia HA 2 2 cc cavity Microphone Source: Dillon (2001): Hearing Aids HA 2

Real-ear to coupler difference Figure 4. 4 RECD: SPL generated in the average adult

Real-ear to coupler difference Figure 4. 4 RECD: SPL generated in the average adult real ear canal minus SPL generated in an HA 1 2 -cc coupler. Source: Dillon (2001): Hearing Aids

2 -cc coupler and control microphone Photo removed to minimize file space Figure 4.

2 -cc coupler and control microphone Photo removed to minimize file space Figure 4. 5 A hearing aid connected to a coupler, with a control microphone positioned next to the hearing aid microphone. Source: Dillon (2001): Hearing Aids

Gain-frequency response Figure 4. 6 Gain-frequency response (measured with a 60 d. B SPL

Gain-frequency response Figure 4. 6 Gain-frequency response (measured with a 60 d. B SPL input level) and OSPL 90 -frequency response of a BTE measured in a 2 -cc coupler with a swept pure tone. The 60 d. B curve can be read against either axis; the OSPL 90 curve must be read against the left hand axis. Source: Dillon (2001): Hearing Aids

Input-output diagram Figure 4. 7 Input-output diagram of a compression hearing aid at 2

Input-output diagram Figure 4. 7 Input-output diagram of a compression hearing aid at 2 k. Hz (bold line) and lines of constant gain (dotted lines). Source: Dillon (2001): Hearing Aids

Equivalent input noise Figure 4. 8 Equivalent 1/3 -octave input noise of a typical

Equivalent input noise Figure 4. 8 Equivalent 1/3 -octave input noise of a typical hearing aid as a function of frequency, and maximum acceptable 1/3 -octave noise. Source: Dillon (2001): Hearing Aids

REAG = A - C M A F C Figure 4. 9 Location of

REAG = A - C M A F C Figure 4. 9 Location of SPLs involved in the measurement of real-ear aided gain. F is located in the undisturbed sound field (e. g. with the head absent), C is at the control microphone location on the surface of the head, M is at the hearing aid microphone port, and A is within the residual ear canal close to the eardrum. Source: Dillon (2001): Hearing Aids

SPL in ear canal Figure 4. 10 Calculated pattern of SPL in the ear

SPL in ear canal Figure 4. 10 Calculated pattern of SPL in the ear canal versus distance from the eardrum at a frequency of 6 k. Hz. The solid curve is for total reflection from the eardrum with no phase shift at the drum, the dashed line is for 50% power reflected from the drum with no phase shift, and the speckled line is for 50% reflected with a 45 degree phases shift at the drum. Source: Dillon (2001): Hearing Aids

Standing-wave minimum Figure 4. 11 Distance from the eardrum at which SPL in the

Standing-wave minimum Figure 4. 11 Distance from the eardrum at which SPL in the ear canal will be a minimum. Source: Dillon (2001): Hearing Aids

Real-ear aided gain Figure 4. 12 Typical REAG display for a vented, low to

Real-ear aided gain Figure 4. 12 Typical REAG display for a vented, low to medium gain hearing aid, displaying the expected low frequency plateau. Source: Dillon (2001): Hearing Aids

Insertion gain = A - U M U F A F C Unaided C

Insertion gain = A - U M U F A F C Unaided C Aided Figure 4. 13 Location of SPLs involved in the measurement of insertion gain. F is located in the undisturbed sound field (with the head absent), C is at the control microphone location on the surface of the head, M is at the hearing aid microphone port, A is at the eardrum when aided, and U is at the eardrum when unaided. Source: Dillon (2001): Hearing Aids

REIG = REAG - REUG Figure 4. 14 Real ear unaided and aided gains

REIG = REAG - REUG Figure 4. 14 Real ear unaided and aided gains (top half). The Source: Dillon (2001): difference between these curves is the insertion gain, shown as the Hearing Aids shaded region in the top half and as the curve in the lower half.

Probe position for insertion gain (a) (b) (c) (d) Figure 4. 15 Probe positioning

Probe position for insertion gain (a) (b) (c) (d) Figure 4. 15 Probe positioning for measuring insertion gain: (a) noting a landmark on the ear; (b) marking the probe; (c) measuring the unaided response; (d) measuring the aided response. Source: Dillon (2001): Hearing Aids

Calibrating the probe Photo removed to minimize file space Figure 4. 16 Positioning of

Calibrating the probe Photo removed to minimize file space Figure 4. 16 Positioning of the probe microphone against the control microphone during calibration. Source: Dillon (2001): Hearing Aids

Feedback Forward path (gain) Feedback path (attenuation) Figure 4. 17 The feedback mechanism in

Feedback Forward path (gain) Feedback path (attenuation) Figure 4. 17 The feedback mechanism in hearing aids. Source: Dillon (2001): Hearing Aids

Feedback Figure 4. 18 Coupler gain of a hearing aid with the volume control

Feedback Figure 4. 18 Coupler gain of a hearing aid with the volume control adjusted in 2 d. B steps. One further increase resulted in oscillation. Source: Dillon (2001): Hearing Aids

Cross section of earmold Skin around canal Probe-induced feedback path Gap created by probe

Cross section of earmold Skin around canal Probe-induced feedback path Gap created by probe tube Probe tube Figure 4. 19 Leakage paths created by the insertion of a probe tube between an earmold or shell and the ear canal. Source: Dillon (2001): Hearing Aids

Stethoclip Photo removed to minimize file space Figure 4. 20 A stethoclip attached to

Stethoclip Photo removed to minimize file space Figure 4. 20 A stethoclip attached to a CIC hearing aid. Source: Dillon (2001): Hearing Aids

Feedback - ITE Microphone tube detached at either end Wax pushes hearing aid away

Feedback - ITE Microphone tube detached at either end Wax pushes hearing aid away from the canal wall Loose fit of shell Microphone or receiver touching each other or touching case Vent too large, or vent insert fallen out, or vent too close to microphone port, or vent overhung by pinnae Wax directs sound into vent or slit leak Receiver tube detached at either end Figure 4. 21 Common leakage points, leading to feedback oscillation, in ITE, ITC, and CIC hearing aids. Source: Dillon (2001): Hearing Aids

Feedback - BTE Split in earhook Tubing too loose a fit on earhook Wax

Feedback - BTE Split in earhook Tubing too loose a fit on earhook Wax pushes earmold away from the canal wall Earhook too loose a fit on aid Wax directs sound into vent or slit leak Microphone or receiver touching case Tubing split Earmold too loose Vent too large, or vent insert fallen out Figure 4. 22 Common leakage points, leading to feedback oscillation, in BTE hearing aids. Source: Dillon (2001): Hearing Aids