comparison NIKHEFEUDET card and SiemensLake Shore Hall probe
comparison NIKHEF/EUDET card and Siemens/Lake Shore Hall probe F. Bergsma CERN 7/2/2008 Results of last meeting 12/9/2007 Recalibration of cards used in EUDET mapping at DESY NIKHEF card with Lakeshore HGT 2100 EUDET card with Siemens KSY 44 More tests on NIKHEF card with KSY 44 Conclusions
NIKHEF Sensor +10 G |B| 1 -5 G T +10 G |B| -5 G 2 T 0. 37 T 1. 14 T 1. 4 T 9 x θ, 16 x φ, step=22. 5 o NIKHEF after one year
NIKHEF Sensor +10 G |B| 3 -5 G T +10 G |B| -5 G 4 T 0. 37 T 1. 14 T 1. 4 T 9 x θ, 16 x φ, step=22. 5 o NIKHEF after one year
EUDET Sensor new +10 G prototype |B| 1 -5 G T +10 G |B| -5 G 2 T 0. 37 T 1. 14 T 1. 4 T 9 x θ, 16 x φ, step=22. 5 o EUDET Temp from DS 60 Immediat after cal.
EUDET Sensor new +10 G prototype |B| 1 -5 G T +10 G |B| -5 G 2 T 0. 37 T 1. 14 T 1. 4 T 9 x θ, 16 x φ, step=22. 5 o EUDET Temp from DS 60 After 6 days
EUDET Sensor new +10 G prototype |B| 1 -5 G T +10 G |B| -5 G 2 T 0. 37 T 1. 14 T 1. 4 T 9 x θ, 16 x φ, step=22. 5 o EUDET Temp from DS 60 After 0. 24 year
EUDET Sensor new +10 G prototype |B| 3 -5 G T +10 G |B| -5 G 4 T 0. 37 T 1. 14 T 1. 4 T 9 x θ, 16 x φ, step=22. 5 o EUDET Temp from DS 60 Immediat after cal.
EUDET Sensor new +10 G prototype |B| 3 -5 G T +10 G |B| -5 G 4 T 0. 37 T 1. 14 T 1. 4 T 9 x θ, 16 x φ, step=22. 5 o EUDET Temp from DS 60 After 6 days
EUDET Sensor new +10 G prototype |B| 3 -5 G T +10 G |B| -5 G 4 T 0. 37 T 1. 14 T 1. 4 T 9 x θ, 16 x φ, step=22. 5 o EUDET Temp from DS 60 After 0. 24 years
+15 G Eudet cards used at DESY card 5 0 days -5 G +15 G 6 months -5 G + 15 G 6 months recal -5 G 9 x θ, 16 x φ, step=22. 5 o +15 G 6 months +40 days diff -5 G +15 G 12 cards tested, look similar 6 months +40 days total -5 G 0. 37 T 1. 14 T 1. 4 T Measurements at 22 0 C Calibration done at 20, 24 0 C
Eudet cards used at DESY card 6 0 days 6 months recal 9 x θ, 16 x φ, step=22. 5 o 6 months +40 days diff 12 cards tested, look similar 6 months +40 days total 0. 37 T 1. 14 T 1. 4 T Measurements at 22 0 C Calibration done at 20, 24 0 C
Eudet cards used at DESY card 7 0 days 6 months recal 9 x θ, 16 x φ, step=22. 5 o 6 months +40 days diff 12 cards tested, look similar 6 months +40 days total 0. 37 T 1. 14 T 1. 4 T Measurements at 22 0 C Calibration done at 20, 24 0 C
2 x NIKHEF CARD with 3 X Lakeshore HGT 2100
+20 G NIKHEF card with HGT 2100 card 2 After 0 days -5 G +20 G After 7 days -5 G +20 G 5 After 47 days recal at 40 days -5 G +60 G ! After 47 days total -5 G 0. 37 T 1. 14 T 1. 4 T 9 x θ, 16 x φ, step=22. 5 o
+20 G NIKHEF card with HGT 2100 card 3 After 0 days -5 G +20 G After 7 days -5 G +20 G After 47 days recal at 40 days -5 G +60 G ! After 47 days total -5 G 0. 37 T 1. 14 T 1. 4 T 9 x θ, 16 x φ, step=22. 5 o
1 x EUDET card with 3 X Siemens KSY 44
Eudet card with KSY 44 +5 G After 0 days -5 G After 21 days After 34 days 9 x θ, 16 x φ, step=22. 5 o After 41 days After 48 days 0. 37 T 1. 14 T 1. 4 T
Do we have an effect due to excessive heating up during soldering? On the NIKHEF HGT 2100 cards we have soldered on a bare smd, no heat sinking, many tries How does the KSY 44 behave after soldering?
+5 G 2 G 0 days NIKHEF card KSY 44 normal operation card 1 -5 G 28 days 40 days 9 x θ, 16 x φ, step=22. 5 o 47 days 54 days 0. 37 T 1. 14 T 1. 4 T Measurements at 22 0 C Calibration done at 20, 24 0 C
+5 G 2 G 0 days NIKHEF card KSY 44 solder test card 4 -5 G 28 days 0. 37 T 1. 14 T 1. 4 T 40 days Simulation of two times soldering = 2 x 2 sec contact with soldering iron/pin 47 days Phenomenon not seen in ATLAS: Time between soldering and calibration more than one month 54 days After two weeks back to 2 G limit Measurements at 22 0 C Calibration done at 20, 24 0 C
Aging studies for ATLAS didn’t show the solder effect because the sensors were produced at NIKHEF, there was always a long ( > 1 month) delay between soldering and calibration Can the HGT 2100 be used? Important points: short solder time heat sinking wait between soldering and calibration (heat treatment? ) further investigation necessary What is the difference between KSY 44 and HGT 2100? HGT 2100 is claimed to be Ga. As, but shows much smaller planar Hall effect than KSY 44 Is the HGT 2100 similar to the KSY 14 ? Ion implanted vs. MOVPE?
Which Hall probe to choose: Lakeshore HGT 2100 very sensitive to soldering, maybe not usable more tests necessary KSY 44 ok but no longer in production forgotten stock of about 300 ( bought before the 7000 of ATLAS) still available around the world e. g 1000 pcs for 7. 5 euro/pc maybe new production GH-700 FW Bell stock problems under investigation Toshiba THS 119 NIKHEF one B-sensor received to be tested